FDA Adverse Event Injury Summary report: N

ENPULSE

MDR report key: 2130235 · Received June 16, 2011

Report

Report Number
6000094-2011-01069
Event Type
Injury
Date Received
June 16, 2011
Manufacturer
MEDTRONIC S.A.
Product Code
DXY
PMA / PMN Number
P980035/S032
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS ANALYZED REVEALING A MODE CHANGE/ POLARITY SWITCH ON THE VENTRICULAR LEAD FROM BIPOLAR TO UNIPOLAR.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) GAVE A POLARITY SWITCH WARNING AND SWITCHED TO UNIPOLAR WHEN IT WAS PROGRAMMED TO BIPOLAR AND MONITOR ONLY. THE IPG WILL BE REPROGRAMMED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENPULSE IMPLANTABLE PULSE GENERATOR DXY MEDTRONIC S.A. E2DR01 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention