FDA Adverse Event
Malfunction
Summary report: N
PIN COLLET
MDR report key: 2130189
·
Received May 5, 2011
Report
- Report Number
- 1811755-2011-01611
- Event Type
- Malfunction
- Date Received
- May 5, 2011
- Date of Event
- March 18, 2011
- Report Date
- March 18, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- KIJ
- PMA / PMN Number
- K972367
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE EVENT CONFIRMED BASED ON THE EXAMINATION. THE IMPREGLON PROCESS DUPONT COATING WORE OFF THE COLLET. THE COATING WEAR CAN CAUSE A WIRE PIN TO STICK AND NOT HOLD IN THE ATTACHMENT DUE TO INCREASED FRICTION BETWEEN THE WEDGE COLLET AND LARGE COLLET. THE CUSTOMER WAS SENT A REPLACEMENT DEVICE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PIN GETS STUCK IN THE PIN COLLET. THE EVENT DID NOT OCCUR DURING A PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PIN COLLET | KIJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |