FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2130180 · Received June 16, 2011

Report

Report Number
2649622-2011-10127
Event Type
Injury
Date Received
June 16, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
Z-0475-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION.  BASED ON THE INFORMATION RECEIVED AND WITHOUT THE RETURN OF THE PRODUCT, IT COULD NOT BE DETERMINED IF THIS DEVICE PERFORMED AS DESCRIBED IN THE FIELD ACTION.

Description of Event or Problem · 1

IT WAS FURTHER REPORTED THAT T-WAVE OVERSENSING (TWOS) HAS RECURRED; WHEN THE HEART RATE GETS HIGH THERE IS T-WAVE OVERSENSING (TWOS) POST SENSE. SENSING WAS REPROGRAMMED FURTHER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS T-WAVE OVERSENSING ON THE RIGHT VENTRICULAR LEAD. THE DEVICE WAS REPROGRAMMED AND THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 35 YR Required Intervention (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB