FDA Adverse Event
Injury
Summary report: N
SPRINT QUATTRO SECURE
MDR report key: 2130180
·
Received June 16, 2011
Report
- Report Number
- 2649622-2011-10127
- Event Type
- Injury
- Date Received
- June 16, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Removal / Correction Number
- Z-0475-2011
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Additional Manufacturer Narrative · 1
THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION. BASED ON THE INFORMATION RECEIVED AND WITHOUT THE RETURN OF THE PRODUCT, IT COULD NOT BE DETERMINED IF THIS DEVICE PERFORMED AS DESCRIBED IN THE FIELD ACTION.
Description of Event or Problem · 1
IT WAS FURTHER REPORTED THAT T-WAVE OVERSENSING (TWOS) HAS RECURRED; WHEN THE HEART RATE GETS HIGH THERE IS T-WAVE OVERSENSING (TWOS) POST SENSE. SENSING WAS REPROGRAMMED FURTHER.
Description of Event or Problem · 1
IT WAS REPORTED THAT THERE WAS T-WAVE OVERSENSING ON THE RIGHT VENTRICULAR LEAD. THE DEVICE WAS REPROGRAMMED AND THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 35 YR | Required Intervention | (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB |