FDA Adverse Event Malfunction Summary report: N

CARELINK

MDR report key: 2130162 · Received June 16, 2011

Report

Report Number
2183613-2011-00191
Event Type
Malfunction
Date Received
June 16, 2011
Manufacturer
MEDTRONIC MILACA INC.
Product Code
NVZ
PMA / PMN Number
P890003/S102
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): ANALYSIS FOUND THAT THE ASSEMBLY TESTED OUT OF SPECIFICATION ELECTRICALLY DUE TO A SERIOUS COMPONENT BURN.

Description of Event or Problem · 1

IT WAS REPORTED BY THE PATIENT THAT FOLLOWING A POWER OUTAGE THE POWER CORD FOR THE CARELINK MONITOR "BLEW UP INTO PIECES." A REPLACEMENT CORD WAS SENT BUT THE MONITOR WOULD NOT WORK. THE MONITOR WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE THERE WAS A POWER OUTAGE AT THE PATIENT'S HOME AND WHEN THE POWER CAME BACK ON THE POWER CORD CONNECTED TO THE CARELINK MONITOR (CLM) "BLEW UP INTO PIECES". A NEW POWER CORD FOR THE CLM WAS MAILED TO THE PATIENT AND THE MONITOR STILL DID NOT WORK. A NEW CLM WILL BE MAILED TO THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CARELINK PATIENT MONITORING INSTRUMENT NVZ MEDTRONIC MILACA INC. 2490C ASKU

Patients

Seq Age Sex Outcome Treatment
1 79 YR Other 6945 IMPLANTABLE TACHY LEAD| 5076 IMPLANTABLE PACING LEAD| C154DWK IMPLANTABLE PACEMAKER/CARDIO/DEFIB| POWER CORD| 4193 IMPLANTABLE PACING LEAD| C154DWK IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 6945 IMPLANTABLE TACHY LEAD| 5076 IMPLANTABLE PACING LEAD| 4193 IMPLANTABLE PACING LEAD| POWER CORD| C154DWK IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 6945 IMPLANTABLE TACHY LEAD| 5076 IMPLANTABLE PACING LEAD| 4193 IMPLANTABLE PACING LEAD