SIGMA 100 VVI
Report
- Report Number
- 6000094-2011-01065
- Event Type
- Injury
- Date Received
- June 16, 2011
- Date of Event
- March 25, 2011
- Manufacturer
- MEDTRONIC S.A.
- Product Code
- DXY
- PMA / PMN Number
- P980035/S2
- Removal / Correction Number
- Z-1510/1519-2009
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS.
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) PRELIMINARY TESTING REVEALED NO OUTPUT AND NO TELEMETRY. THE NO OUTPUT AND NO TELEMETRY CONDITIONS WERE THE RESULT OF LIFTED HYBRID BOND WIRES.
.
IT WAS REPORTED THAT ALL PACEMAKER FUNCTIONS HAVE FAILED. THERE WAS NO TELEMETRY OR OUTPUT. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SIGMA 100 VVI | IMPLANTABLE PULSE GENERATOR | DXY | MEDTRONIC S.A. | SVVI103 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |