MAXIMO II CRT-D
Report
- Report Number
- 6000144-2011-03304
- Event Type
- Injury
- Date Received
- June 16, 2011
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- NIK
- PMA / PMN Number
- P010031/S084
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): ACTUAL LONGEVITY IS < 80% OF 99.9% LONGEVITY LIMIT. THE DEVICE WAS FULLY FUNCTIONAL, WITH NO HIGH CURRENT DRAIN OR EVIDENCE OF BATTERY PROBLEMS. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.
IT WAS REPORTED THAT THE BATTERY LONGEVITY WAS LESS THAN FOUR YEARS. THE DEVICE WAS EXPLANTED AND REPLACED. IT WAS ALSO REPORTED THAT THE LEFT VENTRICULAR (LV) OUTPUT WAS HIGH. ONE OF THE LV LEADS IS NO LONGER ACTIVE AND THE OTHER IS STILL IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MAXIMO II CRT-D | IMPLANTABLE PACEMAKER/CARDIO/DEFIB | NIK | MEDTRONIC MED REL, INC. | D284TRK | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 78 YR | Hospitalization| O| R | 4076 IMPLANTABLE PACING LEAD| 6947 IMPLANTABLE TACHY LEAD |