SIENRHYTHM DR
Report
- Report Number
- 6000094-2011-01058
- Event Type
- Injury
- Date Received
- June 16, 2011
- Manufacturer
- MEDTRONIC S.A.
- Product Code
- DXY
- PMA / PMN Number
- P980035/S38
- Removal / Correction Number
- Z-1440-2010
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HK
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS.
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY (B)(4): THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS ANALYZED AND REVEALED THE FOLLOWING: IMPEDANCE - HIGH RESISTANCE/IMPEDANCE AND HIGH BATTERY IMPEDANCE LEAD TO ELECTIVE REPLACEMENT INDICATOR (ERI). BATTERY VOLTAGE - BATTERY DEPLETION INDICATED/ERI AND THE ERI WAS DUE TO HIGH BATTERY IMPEDANCE LOGGED ON (B)(6) 2010, PRIOR TO EXPLANT. RAMWARE VERSION 8 WAS INSTALLED ON (B)(6) 2010.
IT WAS REPORTED THAT THE DEVICE HAD A SHARP DROP IN BATTERY VOLTAGE AND POSSIBLY REACHED ELECTIVE REPLACEMENT INDICATOR (ERI) EARLY. THE DEVICE WILL BE EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SIENRHYTHM DR | IMPLANTABLE PULSE GENERATOR | DXY | MEDTRONIC S.A. | P1501DR | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |