FDA Adverse Event Injury Summary report: N

SIENRHYTHM DR

MDR report key: 2130063 · Received June 16, 2011

Report

Report Number
6000094-2011-01058
Event Type
Injury
Date Received
June 16, 2011
Manufacturer
MEDTRONIC S.A.
Product Code
DXY
PMA / PMN Number
P980035/S38
Removal / Correction Number
Z-1440-2010
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HK
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY (B)(4): THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS ANALYZED AND REVEALED THE FOLLOWING: IMPEDANCE - HIGH RESISTANCE/IMPEDANCE AND HIGH BATTERY IMPEDANCE LEAD TO ELECTIVE REPLACEMENT INDICATOR (ERI). BATTERY VOLTAGE - BATTERY DEPLETION INDICATED/ERI AND THE ERI WAS DUE TO HIGH BATTERY IMPEDANCE LOGGED ON (B)(6) 2010, PRIOR TO EXPLANT. RAMWARE VERSION 8 WAS INSTALLED ON (B)(6) 2010.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD A SHARP DROP IN BATTERY VOLTAGE AND POSSIBLY REACHED ELECTIVE REPLACEMENT INDICATOR (ERI) EARLY. THE DEVICE WILL BE EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SIENRHYTHM DR IMPLANTABLE PULSE GENERATOR DXY MEDTRONIC S.A. P1501DR ASKU

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R