FDA Adverse Event Malfunction Summary report: N

NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE

MDR report key: 2130048 · Received May 5, 2011

Report

Report Number
1811755-2011-01612
Event Type
Malfunction
Date Received
May 5, 2011
Date of Event
March 4, 2011
Report Date
March 4, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
FYD
PMA / PMN Number
K012991
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS EVALUATED BY A FIELD SERVICE REPRESENTATIVE. IT WAS FOUND THAT THE A/C PLUG WAS PARTIALLY MELTED INSIDE. THE A/C PLUG WAS REPLACED AND RETURNED TO SERVICE AFTER PASSING ALL TESTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE POWER PLUG WAS PARTIALLY MELTED. ADDITIONAL INFORMATION IS BEING REQUESTED FROM THE ACCOUNT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE FYD STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK