- Report Number
- 1811755-2011-01612
- Event Type
- Malfunction
- Date Received
- May 5, 2011
- Date of Event
- March 4, 2011
- Report Date
- March 4, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- FYD
- PMA / PMN Number
- K012991
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS EVALUATED BY A FIELD SERVICE REPRESENTATIVE. IT WAS FOUND THAT THE A/C PLUG WAS PARTIALLY MELTED INSIDE. THE A/C PLUG WAS REPLACED AND RETURNED TO SERVICE AFTER PASSING ALL TESTS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE POWER PLUG WAS PARTIALLY MELTED. ADDITIONAL INFORMATION IS BEING REQUESTED FROM THE ACCOUNT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE | FYD | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |