FDA Adverse Event Malfunction Summary report: N

MAESTRO DRILL WITH HANDSWITCH

MDR report key: 2130044 · Received May 5, 2011

Report

Report Number
1811755-2011-01531
Event Type
Malfunction
Date Received
May 5, 2011
Date of Event
April 6, 2011
Report Date
April 7, 2011
Manufacturer
STRYKER INSTRUMENT KALAMAZOO
Product Code
HBB
PMA / PMN Number
K041754
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HANDPIECE WAS RECEIVED BY THE MANUFACTURER, AND THE COMPLAINT WAS CONFIRMED. BASED ON THE QUALITY INVESTIGATION, THE DRILL'S HANDSWITCH ASSEMBLY BECAME DISPLACED, WHICH DISABLED THE SAFETY MODE AND ALLOWED THE DRILL TO OPERATE WHEN NOT INTENDED. IT IS UNK HOW THE HANDSWITCH ASSEMBLY BECAME DISPLACED. THE HANDSWITCH WAS REPAIRED AND ADDITIONAL PREVENTATIVE MAINTENANCE WAS ALSO PERFORMED. THE DRILL WAS RETURNED TO THE USER FACILITY.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER, THE DRILL ACTIVATED WHILE IN SAFE MODE. THIS EVENT DID NOT OCCUR IN A SURGICAL ENVIRONMENT, SO THERE WAS NO PT INVOLVEMENT. NO USER INJURY WAS REPORTED, AND NO OTHER ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MAESTRO DRILL WITH HANDSWITCH HBB STRYKER INSTRUMENT KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK