FDA Adverse Event Malfunction Summary report: N

MAESTRO DRILL WITH HANDSWITCH

MDR report key: 2130041 · Received May 5, 2011

Report

Report Number
1811755-2011-01538
Event Type
Malfunction
Date Received
May 5, 2011
Date of Event
March 25, 2011
Report Date
April 11, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
HBB
PMA / PMN Number
K041754
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DRILL WAS RECEIVED BY THE MANUFACTURER, AND THE COMPLAINT WAS CONFIRMED. INTERNAL COMPONENTS WERE EVALUATED AND AN O-RING WAS FOUND TO BE DISPLACED. PREVENTING A VALVE FROM SEATING PROPERLY AND STOPPING THE PNEUMATIC AIR FLOW TO THE DRILL. THIS ISSUE WILL RESULT IN THE DRILL CONTINUING TO OPERATE, EVEN WHEN THE TRIGGER IS NOT ACTIVATED. THE DRILL WAS REPAIRED AND ADDITIONAL PREVENTATIVE MAINTENANCE WAS ALSO PERFORMED. THE DEVICE WAS RETURNED TO THE USER FACILITY.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER, THE DRILL OPERATED WHEN THE TRIGGER WAS NOT ACTIVATED. THIS EVENT DID NOT OCCUR IN A SURGICAL ENVIRONMENT, SO THERE WAS NO PT INVOLVEMENT. NO USER INJURY WAS REPORTED, AND NO OTHER ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MAESTRO DRILL WITH HANDSWITCH HBB STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK