FDA Adverse Event Malfunction Summary report: N

NEPTUNE 2 DOCKING STATION (120V)

MDR report key: 2130040 · Received May 5, 2011

Report

Report Number
1811755-2011-01599
Event Type
Malfunction
Date Received
May 5, 2011
Date of Event
March 15, 2011
Report Date
March 15, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
FYD
PMA / PMN Number
K012991
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS EVALUATED BY A FIELD SERVICE TECHNICIAN AND THERE WERE NO LEAKS FOUND ON THE DOCKER. THERE WAS A WATER LEAK FOUND FROM THE BACK FLOW PREVENTER AND WATER HAMMER SYSTEM. THE CUSTOMER WAS INFORMED THAT THE ISSUE WAS A HOSPITAL PLUMBING ISSUE.

Description of Event or Problem · 1

IT WAS REPORTED THAT SOAPY WATER WAS LEAKING THROUGH TO THE FLOOR BELOW. THERE WAS NO PT INVOLVEMENT OR ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NEPTUNE 2 DOCKING STATION (120V) FYD STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK