FDA Adverse Event Malfunction Summary report: N

DAVINCI XI

MDR report key: 21289360 · Received February 3, 2025

Report

Report Number
2955842-2025-01061
Event Type
Malfunction
Date Received
February 3, 2025
Date of Event
January 13, 2025
Report Date
January 13, 2025
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874110898
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. FSE REPLACED INTEGRATED ELECTROSURGICAL UNIT (IESU) TO RESOLVE THE ISSUE. THE SYSTEM WAS VERIFIED AND READY TO USE. THE UNIT WAS RETURNED FOR FAILURE ANALYSIS, AND THE REPORTED FAILURE (C-34/C-00 ERRORS) WAS CONFIRMED AND REPRODUCED ON GENERATOR CONNECTED TO SYSTEM. VISUAL INSPECTION SHOWED BEZEL WAS CRACKED TOP LEFT CORNER. THE COVER HAD A DEEP SCRATCH TO THE METAL ON SIDE. THERE WAS C-34 ERROR DURING START-UP AND MONO COAG ACTIVATION UNIT THAT WAS PLACED ON SYSTEM AND WAS RUN IN NORMAL MODE. THE ROOT CAUSE WAS ATTRIBUTED TO A HIGH FREQUENCY VOLTAGE ISSUE.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED INGUINAL HERNIA SURGICAL PROCEDURE, THERE WERE REPEATED ERRORS ON THE INTEGRATED ELECTROSURGICAL UNIT (IESU). THE CUSTOMER CLARIFIED THAT THEY OBSERVED ERROR C-34. THE TECHNICAL SERVICE ENGINEER (TSE) CONFIRMED ERROR C-34 ALONG WITH M-11, M-18, AND C-06. THE CUSTOMER ATTEMPTED TO USE BOTH MONOPOLAR ENERGY PORTS AND REBOOTED THE IESU MULTIPLE TIMES WITH NO RESOLUTION. THE CUSTOMER DID NOT HAVE AN AVAILABLE BACKUP GENERATOR AND OPTED TO CONTINUE THE CASE USING THE E-100 GENERATOR WITH A VESSEL SEALER INSTRUMENT. THE PROCEDURE WAS COMPLETING AS PLANNED WITH NO REPORTED INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2088525 DAVINCI XI VISION SIDE SYSTEM NAY INTUITIVE SURGICAL, INC 381121-41 N/A 00886874110898

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown DA VINCI INSTRUMENTS AND ACCESSORIES