FDA Adverse Event Malfunction Summary report: N

GSM ANTENNA

MDR report key: 21285651 · Received January 31, 2025

Report

Report Number
3004936110-2025-00181
Event Type
Malfunction
Date Received
January 31, 2025
Date of Event
October 16, 2024
Report Date
January 31, 2025
Manufacturer
ABBOTT MEDICAL.
Product Code
MOM
PMA / PMN Number
P100045
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

VISUAL INSPECTION OF RETURNED MATERIAL REVEALED FUNCTIONAL DAMAGE TO THE GSM ANTENNA. DAMAGE TO THE GSM ANTENNA CAME IN THE FORM OF THE RADOME PORTION SPLIT OPEN WITH THE POTENTIAL FOR EXPOSED WIRING.

Description of Event or Problem · 0

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS CONFIRMING A BREAK IN THE GSM ADAPTOR RESULTING IN THE POTENTIAL FOR EXPOSED WIRES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
494401 GSM ANTENNA System, hemodynamic, implantable MOM ABBOTT MEDICAL. CM1040

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown