FDA Adverse Event Malfunction Summary report: N

CARESCAPE

MDR report key: 21285352 · Received January 31, 2025

Report

Report Number
2112667-2025-00651
Event Type
Malfunction
Date Received
January 31, 2025
Date of Event
January 6, 2025
Report Date
January 31, 2025
Manufacturer
DATEX-OHMEDA, INC.
Product Code
CBK
UDI-DI
00840682102346
PMA / PMN Number
K210384
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DISTRIBUTOR PERFORMED A CHECKOUT OF THE EQUIPMENT AND CONFIRMED THE REPORTED COMPLAINT. THE ELECTRONIC FLOW CONTROL (EFC) VALVE WAS CALIBRATED TO RESOLVE THE ISSUE. BLOCK A: NO REPORT OF PATIENT INVOLVEMENT. LEGAL MANUFACTURER: HCS MADISON - 3030 OHMEDA DR, USA MADISON, WI 53718.

Description of Event or Problem · 0

IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN INSUFFICIENT OXYGEN OR FRESH GAS FLOW. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
462870 CARESCAPE VENTILATOR, CONTINUOUS, FACILITY USE CBK DATEX-OHMEDA, INC. R860 NA 00840682102346

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown