FDA Adverse Event Malfunction Summary report: N

CERENOVUS ENTERPRISE

MDR report key: 21284712 · Received January 31, 2025

Report

Report Number
3008114965-2025-00071
Event Type
Malfunction
Date Received
January 31, 2025
Date of Event
January 17, 2025
Report Date
April 21, 2025
Manufacturer
MEDOS INTERNATIONAL SARL
Product Code
NJE
UDI-DI
10886704075288
PMA / PMN Number
H60001
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MANUFACTURER¿S REF. NO: (B)(4). INFORMATION REGARDING PATIENT IDENTIFIER, DATE OF BIRTH, AGE, SEX, GENDER, WEIGHT, RACE, AND ETHNICITY WERE NOT PROVIDED. SECTION E.1: THE INITIAL REPORTER PHONE: (B)(6). SECTION H.3: THE DEVICE IS AVAILABLE TO BE RETURNED FOR EVALUATION AND TESTING. HOWEVER, IT HAS NOT BEEN RECEIVED TO DATE. IF THE DEVICE RETURNS, A DEVICE INVESTIGATION WILL BE PERFORMED. LAKE REGION MEDICAL PERFORMED A REVIEW OF THE DEVICE HISTORY RECORDS RELATIVE TO THE MANUFACTURING, INSPECTION, AND PACKAGING OF THE LOT 9061789. THE HISTORY RECORD INDICATES THIS PRODUCT WAS FINAL INSPECTION TESTED AT LAKE REGION MEDICAL AND WAS DETERMINED TO BE ACCEPTABLE. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY JOHNSON & JOHNSON MEDTECH, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, JOHNSON & JOHNSON MEDTECH, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. MISSING INFORMATION FROM THIS REPORT IS IDENTIFIED AS BLANK; THIS INFORMATION WAS NOT PROVIDED IN THE REPORTED EVENT OR AVAILABLE AT THE TIME OF REPORT SUBMISSION. THE MANUFACTURER WILL SUBMIT A SUPPLEMENTAL REPORT IF NEW FACTS ARISE WHICH MATERIALLY ALTER INFORMATION SUBMITTED IN A PREVIOUS MDR REPORT. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN 30 DAYS OF RECEIPT.

Additional Manufacturer Narrative · 0

MANUFACTURER¿S REF. NO: (B)(4). THE PURPOSE OF THIS MDR SUBMISSION IS TO REPORT THE INVESTIGATIONAL FINDING OF THE RETURNED DEVICE. THE REPORT WILL ALSO INCLUDE A CORRECTION TO THE PRIMARY UDI NUMBER. THE COMPLAINT PRODUCT WAS RETURNED AND RECEIVED FOR EVALUATION AND ANALYSIS. THE INVESTIGATION IS DOCUMENTED BELOW. INVESTIGATION SUMMARY: A NON-STERILE 4MM X 23MM NO TIP ENTERPRISE¿ 2 VASCULAR RECONSTRUCTION DEVICE WAS RECEIVED CONTAINED IN THE DECONTAMINATION POUCH. VISUAL INSPECTION WAS PERFORMED. IT WAS NOTED THAT ONLY THE DETACHED STENT COMPONENT WAS RETURNED FOR EVALUATION. MICROSCOPIC INSPECTION WAS PERFORMED ON THE STENT COMPONENT. IT WAS OBSERVED TO BE IN GOOD CONDITION. THERE WAS NO STRUCTURAL DAMAGE (I.E., NO BROKEN STRUTS, NO KINKS). IT WAS ALSO NOTED TO BE FULLY EXPANDED WITH BOTH ENDS COMPLETELY FLARED. THE ISSUE REGARDING A STENT BEING IMPEDED IN THE DISTAL END OF THE CONCOMITANT MICROCATHETER CANNOT BE EVALUATED THROUGH FUNCTIONAL TESTING. THE STENT MUST BE INSIDE THE INTRODUCER TUBE TO PERFORM THE FUNCTIONAL ANALYSIS. ADDITIONALLY, THE RETURNED COMPONENT DID NOT PRESENT DAMAGES THAT SUGGEST THAT IT WAS FORCIBLY ADVANCED. WITH THE LIMITED INFORMATION AVAILABLE, A CONCLUSIVE CAUSE CANNOT BE DETERMINED; HOWEVER, IT IS POSSIBLE THAT CLINICAL AND PROCEDURAL FACTORS, INCLUDING DEVICE MANIPULATION AND OPERATOR'S TECHNIQUE, MAY HAVE CONTRIBUTED TO THE REPORTED FAILURE. AT THIS TIME, THERE IS NO EVIDENCE TO SUPPORT THAT THE ISSUE REPORTED IN THE COMPLAINT IS A RESULT OF A DEFECT INHERENTLY RELATED TO THE DEVICE. THE DELIVERY WIRE AND INTRODUCER COMPONENTS MIGHT HAVE GOTTEN LOST SOMETIME DURING THE POST-OPERATIVE HANDLING OF THE DEVICE. IF ADDITIONAL INFORMATION OR COMPONENTS ARE RECEIVED AT A LATER TIME, THIS INVESTIGATION WILL BE REASSESSED ACCORDINGLY. THE STENT DETACHMENT WAS NOT ORIGINALLY REPORTED IN THE COMPLAINT AND IS NOT CONSIDERED RELATED TO THE ENCOUNTERED ISSUE. LAKE REGION MEDICAL PERFORMED A REVIEW OF THE DEVICE HISTORY RECORDS RELATIVE TO THE MANUFACTURING, INSPECTION, AND PACKAGING OF THE LOT 9061789. THE HISTORY RECORD INDICATES THIS PRODUCT WAS FINAL INSPECTION TESTED AT LAKE REGION MEDICAL AND WAS DETERMINED TO BE ACCEPTABLE. AS PART OF JOHNSON & JOHNSON MEDTECH QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. IT SHOULD BE NOTED THAT PRODUCT FAILURE COULD BE CAUSED BY MULTIPLE FACTORS. THE INSTRUCTIONS FOR USE (IFU) DOES CONTAIN THE FOLLOWING RECOMMENDATIONS: IF RESISTANCE IS MET DURING MANIPULATION, DETERMINE THE CAUSE OF RESISTANCE BEFORE PROCEEDING. DO NOT APPLY UNDUE FORCE IF RESISTANCE IS ENCOUNTERED AT ANY POINT DURING STENT MANIPULATION. WITHDRAW THE UNIT AND ADVANCE A NEW ONE. BASED ON THE MANUFACTURING DOCUMENTATION REVIEW, THERE IS NO INDICATION THAT THE EVENT IS RELATED TO THE DEVICE MANUFACTURING PROCESS. AS PART OF THE POST MARKET SURVEILLANCE PROGRAM, INFORMATION FROM THIS COMPLAINT IS TRENDED FOR STATISTICAL SIGNALS AND CORRECTIVE / PREVENTIVE ACTION MAY BE TRIGGERED AT A LATER TIME. SINCE THERE WAS NO EVIDENCE TO SUGGEST THE EVENT WAS RELATED TO A MANUFACTURING OR DESIGN ISSUE, NO CORRECTIVE ACTIONS WILL BE TAKEN AT THIS TIME. CORRECTED SECTION: D.4: PRIMARY UDI NUMBER. UPDATED SECTIONS: B.4, D.4, G.3, G.6. H.2, H.3, H.6, AND H.11. THE MANUFACTURER WILL SUBMIT A SUPPLEMENTAL REPORT IF NEW FACTS ARISE WHICH MATERIALLY ALTER INFORMATION SUBMITTED IN A PREVIOUS MDR REPORT. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN 30 DAYS OF RECEIPT.

Additional Manufacturer Narrative · 0

MANUFACTURER¿S REF. NO: (B)(4). THE PURPOSE OF THIS MDR SUBMISSION IS TO REPORT THAT THE PRODUCT ANALYSIS LAB RECEIVED THE COMPLAINT DEVICE ON 26-FEB-2025. A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED ONCE THE PRODUCT INVESTIGATION HAS BEEN COMPLETED. UPDATED SECTIONS. THE MANUFACTURER WILL SUBMIT A SUPPLEMENTAL REPORT IF NEW FACTS ARISE WHICH MATERIALLY ALTER INFORMATION SUBMITTED IN A PREVIOUS MDR REPORT. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN 30 DAYS OF RECEIPT.

Description of Event or Problem · 0

THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION PROCEDURE, THE 4MM X 23MM NO TIP ENTERPRISE¿ 2 VASCULAR RECONSTRUCTION DEVICE (ENCR402300 / 9061789) WAS IMPEDED IN THE DISTAL END OF THE CONCOMITANT 150CM X 5CM PROWLER SELECT PLUS MICROCATHETER (606S255X / 31411538) AND COULD NOT PASS THROUGH THE MICROCATHETER. THE PHYSICIAN REMOVED THE STENT AND THE MICROCATHETER FROM THE PATIENT AND REPLACED BOTH DEVICES TO COMPLETE THE PROCEDURE, WHICH WAS REPORTEDLY PROLONGED BY APPROXIMATELY 10 MINUTES. THERE WAS NO REPORT OF ANY NEGATIVE PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
519532 CERENOVUS ENTERPRISE INTRACRANIAL NEUROVASCULAR STENT NJE MEDOS INTERNATIONAL SARL 9061789 10886704075288

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown PROWLER SELECT PLUS 150/5CM.