FDA Adverse Event Malfunction Summary report: N

PROWLER SELECT

MDR report key: 21284708 · Received January 31, 2025

Report

Report Number
3008114965-2025-00072
Event Type
Malfunction
Date Received
January 31, 2025
Date of Event
January 17, 2025
Report Date
April 21, 2025
Manufacturer
MEDOS INTERNATIONAL SARL
Product Code
KRA
UDI-DI
10886704028888
PMA / PMN Number
K021591
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MANUFACTURER¿S REF. NO: (B)(4). INFORMATION REGARDING PATIENT IDENTIFIER, DATE OF BIRTH, AGE, SEX, GENDER, WEIGHT, RACE, AND ETHNICITY WERE NOT PROVIDED. SECTION E.1: THE INITIAL REPORTER PHONE: (B)(6). SECTION H.3: THE DEVICE IS AVAILABLE TO BE RETURNED FOR EVALUATION AND TESTING. HOWEVER, IT HAS NOT BEEN RECEIVED TO DATE. IF THE DEVICE RETURNS, A DEVICE INVESTIGATION WILL BE PERFORMED. A REVIEW OF MANUFACTURING DOCUMENTATION ASSOCIATED WITH THIS LOT (31411538) PRESENTED NO ISSUES DURING THE MANUFACTURING OR INSPECTION PROCESSES RELATED TO THE REPORTED COMPLAINT. THERE WERE NO INTERNAL ACTIONS RELATED TO DEVICE MANUFACTURE OR INSPECTION. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY JOHNSON & JOHNSON MEDTECH, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, JOHNSON & JOHNSON MEDTECH, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. MISSING INFORMATION FROM THIS REPORT IS IDENTIFIED AS BLANK; THIS INFORMATION WAS NOT PROVIDED IN THE REPORTED EVENT OR AVAILABLE AT THE TIME OF REPORT SUBMISSION. THE MANUFACTURER WILL SUBMIT A SUPPLEMENTAL REPORT IF NEW FACTS ARISE WHICH MATERIALLY ALTER INFORMATION SUBMITTED IN A PREVIOUS MDR REPORT. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN 30 DAYS OF RECEIPT.

Additional Manufacturer Narrative · 0

MANUFACTURER¿S REF. NO: PC-(B)(4). THE PURPOSE OF THIS MDR SUBMISSION IS TO REPORT THAT THE PRODUCT ANALYSIS LAB RECEIVED THE COMPLAINT DEVICE ON (B)(6) 2025. A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED ONCE THE PRODUCT INVESTIGATION HAS BEEN COMPLETED. UPDATED SECTIONS: B.4, D.9, G.3, G.6. H.2, H.3, H.6, AND H.11. THE MANUFACTURER WILL SUBMIT A SUPPLEMENTAL REPORT IF NEW FACTS ARISE WHICH MATERIALLY ALTER INFORMATION SUBMITTED IN A PREVIOUS MDR REPORT. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN 30 DAYS OF RECEIPT.

Additional Manufacturer Narrative · 0

MANUFACTURER¿S REF. NO: (B)(4). THE PURPOSE OF THIS MDR SUBMISSION IS TO REPORT THE INVESTIGATIONAL FINDING OF THE RETURNED DEVICE. THE REPORT WILL ALSO INCLUDE A CORRECTION TO THE PRIMARY UDI NUMBER. THE COMPLAINT PRODUCT WAS RETURNED AND RECEIVED FOR EVALUATION AND ANALYSIS. THE INVESTIGATION IS DOCUMENTED BELOW. INVESTIGATION SUMMARY: A NON-STERILE 150CM X 5CM PROWLER SELECT PLUS MICROCATHETER WAS RECEIVED CONTAINED IN THE DECONTAMINATION POUCH. VISUAL INSPECTION WAS PERFORMED. NO APPEARANCE OF DAMAGES WAS OBSERVED. THE MICROCATHETER WAS FLUSHED USING A LAB SAMPLE SYRINGE. AFTER FLUSHING, A LAB SAMPLE GUIDEWIRE WAS INTRODUCED INTO THE MICROCATHETER, AND IT WAS ABLE TO BE ADVANCED THROUGH THE ENTIRE LENGTH OF THE MICROCATHETER WITHOUT NOTICEABLE RESISTANCE. THE INNER DIAMETER (ID) AND OUTER DIAMETER (OD) WERE CONFIRMED TO BE WITHIN SPECIFICATIONS. THE LAB SAMPLE GUIDE WIRE WAS ABLE TO PASS THROUGH THE ENTIRE LENGTH OF THE GUIDE CATHETER WITHOUT SIGNIFICANT RESISTANCE. THE ISSUE DOCUMENTED IN THE COMPLAINT CANNOT BE CONFIRMED WITH THE EVIDENCE AVAILABLE. IT IS POSSIBLE THAT OTHER CLINICAL AND PROCEDURAL FACTORS THAT CANNOT BE REPLICATED DURING THE ANALYSIS MAY HAVE CONTRIBUTED TO THE REPORTED FAILURE. ADDITIONALLY, NO OTHER DAMAGES WERE FOUND ON THE MICROCATHETER THAT COULD HAVE CONTRIBUTED TO THE ISSUE REPORTED DURING THE PROCEDURE. A REVIEW OF MANUFACTURING DOCUMENTATION ASSOCIATED WITH THIS LOT (31411538) PRESENTED NO ISSUES DURING THE MANUFACTURING OR INSPECTION PROCESSES RELATED TO THE REPORTED COMPLAINT. THERE WERE NO INTERNAL ACTIONS RELATED TO DEVICE MANUFACTURE OR INSPECTION. AS PART OF JOHNSON & JOHNSON MEDTECH QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. THE INSTRUCTIONS FOR USE (IFU) CONTAINS THE FOLLOWING CAUTION: ¿ IF STRONG RESISTANCE IS MET DURING MANIPULATION, DISCONTINUE THE PROCEDURE AND DETERMINE THE CAUSE OF RESISTANCE BEFORE PROCEEDING. IF THE CAUSE OF RESISTANCE CANNOT BE DETERMINED, WITHDRAW THE CATHETER AND GUIDEWIRE AS A SYSTEM. BASED ON THE MANUFACTURING DOCUMENTATION REVIEW, THERE IS NO INDICATION THAT THE EVENT IS RELATED TO THE DEVICE MANUFACTURING PROCESS. AS PART OF THE POST MARKET SURVEILLANCE PROGRAM, INFORMATION FROM THIS COMPLAINT IS TRENDED FOR STATISTICAL SIGNALS AND CORRECTIVE / PREVENTIVE ACTION MAY BE TRIGGERED AT A LATER TIME. SINCE THERE WAS NO EVIDENCE TO SUGGEST THE EVENT WAS RELATED TO A MANUFACTURING OR DESIGN ISSUE, NO CORRECTIVE ACTIONS WILL BE TAKEN AT THIS TIME. CORRECTED SECTION: D.4: PRIMARY UDI NUMBER. UPDATED SECTIONS: B.4, D.4, G.3, G.6. H.2, H.3, H.6, AND H.11. THE MANUFACTURER WILL SUBMIT A SUPPLEMENTAL REPORT IF NEW FACTS ARISE WHICH MATERIALLY ALTER INFORMATION SUBMITTED IN A PREVIOUS MDR REPORT. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN 30 DAYS OF RECEIPT.

Description of Event or Problem · 0

THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION PROCEDURE, THE 4MM X 23MM NO TIP ENTERPRISE¿ 2 VASCULAR RECONSTRUCTION DEVICE (ENCR402300 / 9061789) WAS IMPEDED IN THE DISTAL END OF THE CONCOMITANT 150CM X 5CM PROWLER SELECT PLUS MICROCATHETER (606S255X / 31411538) AND COULD NOT PASS THROUGH THE MICROCATHETER. THE PHYSICIAN REMOVED THE STENT AND THE MICROCATHETER FROM THE PATIENT AND REPLACED BOTH DEVICES TO COMPLETE THE PROCEDURE, WHICH WAS REPORTEDLY PROLONGED BY APPROXIMATELY 10 MINUTES. THERE WAS NO REPORT OF ANY NEGATIVE PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
519528 PROWLER SELECT CATHETER, CONTINUOUS FLUSH KRA MEDOS INTERNATIONAL SARL 31411538 10886704028888

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown ENTERPRISE2 4MMX23MM NO TIP.