FDA Adverse Event Malfunction Summary report: Y

QUANTUM ROLLER PUMP 6"

MDR report key: 21282907 · Received January 31, 2025

Report

Report Number
3006073153-2025-00044
Event Type
Malfunction
Date Received
January 31, 2025
Report Date
January 31, 2025
Manufacturer
SPECTRUM MEDICAL LTD
Product Code
DTQ
UDI-DI
05060434420787
PMA / PMN Number
K173834
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DURING THE INVESTIGATION: (B)(4) - FOUND THAT A LID MISS MATCH ERROR OCCURRED CAUSING THE PUMP TO STOP. THIS OCCURS WHEN THE MAGNET IN THE LID OF THE PUMP DOES NOT MATCH UP WITH THE MAGNET IN THE PUMP. USER ARE TRAINED TO OVERRIDE THIS ERROR IN CASES WHEN THIS OCCURS. (B)(4) - INVESTIGATION FOUND A CABLE WITHIN THE PUMP WAS DAMAGED. THE CABLE WAS REPLACED, AND THE PUMP FUNCTIONED AS EXPECTED FROM THAT POINT ONWARDS. (B)(4) - INVESTIGATION FOUND A FAULTY PIN ON THE ENCODER WITHIN THE PUMP WAS DAMAGED. THE ENCODER WAS REPLACED, AND THE PUMP FUNCTIONED AS EXPECTED FROM THAT POINT ONWARDS. PLEASE FIND ATTACHED A SPREADSHEET LISTING THE SERIAL NUMBERS OF THE DEVICES SUMMARISED IN THIS SUMMARY REPORT.

Description of Event or Problem · 0

USERS REPORTED EXPERIENCING PUMP STOP DURING A CASE. IN THESE CASES, THERE WAS NO PATIENT HARM. SPECTRUM MEDICAL CONSIDERS INTERRUPTION TO BLOOD FLOW A REPORTABLE MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
494922 QUANTUM ROLLER PUMP 6" ROLLER PUMP DTQ SPECTRUM MEDICAL LTD QRP6 05060434420787

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown