UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER
Report
- Report Number
- 2050012-2011-02351
- Event Type
- Malfunction
- Date Received
- June 15, 2011
- Date of Event
- May 20, 2011
- Report Date
- May 20, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
CTS DIRECTED CUSTOMER TO TIGHTEN AND SECURE ALL THE FITTINGS AT THE CC SAMPLE AND CC REAGENT PROBES, THE FITTING AT THE CC SAMPLE AND CC REAGENT VALVE BLOCKS, AND THE SYRINGES. TROUBLESHOOTING WITH CTS RESOLVED THE ISSUE FOR THE CUSTOMER. CUSTOMER WAS INSTRUCTED TO MONITOR THE INSTRUMENT AND CALL CTS IF PROBLEM CONTINUED.
CUSTOMER REPORTED TO BECKMAN COULTER INC. (BCI) CUSTOMER TECHNICAL SUPPORT (CTS) OF OBTAINING SUPPRESSED RESULTS FLAGS (BLANK ABSORBANCE LOW, BLANK ABSORBANCE HIGH, OIR LOW, CC CUVETTE ERROR, NO VALUE DETECTED AND FLUID SENSE ERRORS) WHEN RUNNING QC ON THE UNICEL DXC 600 PRO INSTRUMENT. CUSTOMER STATED THAT ONLY THE CARTRIDGE CHEMISTRY SIDE IS AFFECTED, NOT THE MODULAR CHEMISTRY SIDE OF THE INSTRUMENT. CTS DIRECTED CUSTOMER TO CHECK FOR LEAKS ON THE CC SIDE OF THE INSTRUMENT. CUSTOMER FOUND THE CC REAGENT PROBE A DRIPPING AND FLUID POOLING AT THE BOTTOM OF THE CC REAGENT SYRINGE. NO REPORTS OF DEATH OR INJURY HAVE BEEN REPORTED IN CONNECTION WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | DXC 600 PRO | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |