FDA Adverse Event Injury Summary report: N

OT ULTRASMART METER

MDR report key: 2128233 · Received June 15, 2011

Report

Report Number
2939301-2011-04851
Event Type
Injury
Date Received
June 15, 2011
Date of Event
May 22, 2011
Report Date
June 10, 2011
Manufacturer
LIFESCAN INC.
Product Code
NBW
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 ((B)(4) 2011)-DEVICE EVALUATION: THE TEST STRIPS INVOLVED WITH THIS COMPLAINT HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4), 2011 WITH THE FOLLOWING FINDINGS: THE TEST STRIPS HAVE PASSED ALL TESTING WITH NO FAULTS FOUND. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT. 510(K) # IS K021819.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 (07/29/2011) - DEVICE EVALUATION: THE METER INVOLVED WITH THIS COMPLAINT HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE METER HAS PASSED ALL TESTING WITH NO FAULTS FOUND. THE RETAIN TEST STRIPS ALSO PASSED ALL TESTING. A DHR (DEVICE HISTORY RECORD) WAS COMPLETED FOR THIS PRODUCT AND NO DEVIATIONS, NON-CONFORMANCES, OR REWORKS WERE OBSERVED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Description of Event or Problem · 1

ON (B)(6) 2011, THE LAY USER/ PATIENT CONTACTED LIFESCAN (LFS) ALLEGING HER ONETOUCH ULTRASMART METER READ INACCURATELY ERRATIC WHEN COMPARING RESULTS TAKEN ONE AFTER ANOTHER. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE ADVOCATE (CCA) DOCUMENTATION. THE ALLEGED ISSUE BEGAN ON THE MORNING OF (B)(6) 2011. THE PATIENT REPORTED BLOOD GLUCOSE RESULTS OF "130, 151, 159, 103, 95, 172, 149 AND 162 MG/DL" WITH THE SUBJECT METER, PERFORMED WITHIN 20 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS THE EXPECTED VALUE OF <=20% AND/ OR <=20 MG/DL. THE CCA WAS ADVISED THE PATIENT MANAGES HER DIABETES WITH ACTOS PILLS (15 MG). AT THE TIME OF THE ALLEGED ISSUE, THE PATIENT INCREASED HER USUAL DOSE OF MEDICATION. AFTERWARDS, THE PATIENT CLAIMED SHE FELT SYMPTOMS OF SHAKY, "WAS SHIVERING," SWEATY, TINGLING FEELING IN HER HEAD AND DIZZY. THE PATIENT DID NOT SPECIFY RECEIVING MEDICAL TREATMENT. AT THE TIME OF TROUBLESHOOTING, THE CCA NOTED THE METER WAS SET TO THE CORRECT UNIT OF MEASUREMENT AND AN APPROVED SAMPLE SITE WAS USED FOR TESTING. THE PATIENT DID NOT HAVE CONTROL SOLUTION TO TEST THE SUBJECT METER. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT REPORTEDLY DEVELOPED SYMPTOMS SUGGESTIVE OF A SERIOUS INJURY AFTER THE ALLEGED METER ISSUE BEGAN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OT ULTRASMART METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN INC. 3062505

Patients

Seq Age Sex Outcome Treatment
1 49 YR Life Threatening