FDA Adverse Event Injury Summary report: N

MENTOR MEMORYGEL BREAST IMPLANT

MDR report key: 21276842 · Received January 30, 2025

Report

Report Number
1645337-2025-00898
Event Type
Injury
Date Received
January 30, 2025
Date of Event
January 2, 2025
Manufacturer
MENTOR TEXAS
Product Code
FTR
UDI-DI
00081317000341
PMA / PMN Number
P030053
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ON FEBRUARY 18, 2025, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. ON FEBRUARY 20, 2025, THE PRODUCT INVESTIGATION WAS COMPLETED. DEVICE INVESTIGATION SUMMARY: THE PRODUCT WAS RETURNED TO MENTOR FOR EVALUATION. MENTOR CONDUCTED VISUAL INSPECTION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED DEVICE REVEALED THAT THE SM MPP GEL 500CC BREAST IMPLANT HAD A TEAR WITHIN AN AREA OF SHELL ABRASION ON THE POSTERIOR VIEW, MEASURING APPROXIMATELY 1.0 CM. THESE FINDINGS ARE CONSISTENT WITH NORMAL WEAR OF THE DEVICE. SHELL ABRASION SUGGESTS IN-VIVO FOLDING OR CREASING OF THE DEVICE, WHICH MAY BE CAUSED BY CONTINUOUS AND SUSTAINED STRESSES TO THE DEVICE, SUCH AS AN EXCESSIVELY SMALL BREAST POCKET OR THE FOLDING OR WRINKLING OF THE SHELL WITHIN THE BREAST POCKET. IN SOME CASES, BREAST IMPLANTS MAY ALSO EXPERIENCE NORMAL WEAR OVER TIME. AS PART OF MENTOR'S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED ACCORDING TO APPROVED SPECIFICATIONS. BASED ON THE RESULTS OF THIS INVESTIGATION, IT WAS DETERMINED THAT THE PRODUCT MET THE MANUFACTURING RELEASE CRITERIA, AND THAT NORMAL WEAR MAY HAVE CAUSED THE OBSERVED DEFECT. NO CORRECTIVE AND PREVENTIVE ACTION (CAPA) IS REQUIRED NOW.

Additional Manufacturer Narrative · 0

ON FEBRUARY 21, 2025, MENTOR RECEIVED ADDITIONAL INFORMATION INDICATING THAT THE PATIENT'S IMPLANTS WERE REPLACED WITH THE FOLLOWING DEVICES: (RIGHT) 500CC MENTOR MEMORYGEL BREAST IMPLANT CATALOG: 3505001BC LOT: 2011211 SN: (B)(6) AND (LEFT) 500CC MENTOR MEMORYGEL BREAST IMPLANT CATALOG: 3505001BC LOT: 2011211 SN: (B)(6).

Additional Manufacturer Narrative · 0

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: MATERIAL RUPTURE. MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT A PATIENT UNDERWENT PRIMARY BREAST AUGMENTATION WITH TWO 500CC MENTOR MEMORYGEL BREAST IMPLANTS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED, VIA MRI AND MAMMOGRAM, WITH RIGHT BREAST IMPLANT RUPTURE. AS A RESULT, THE PATIENT UNDERWENT BREAST IMPLANT REMOVAL SURGERY ON (B)(6) 2025.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
830211 MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR MENTOR TEXAS 6716623 00081317000341

Patients

Seq Age Sex Outcome Treatment
1 42 YR Female Required Intervention