FDA Adverse Event Malfunction Summary report: N

TPRLC 133 MP TYPE1 PPS SO 16.0

MDR report key: 21272807 · Received January 30, 2025

Report

Report Number
0001825034-2025-00229
Event Type
Malfunction
Date Received
January 30, 2025
Date of Event
January 10, 2025
Report Date
May 30, 2025
Manufacturer
ZIMMER BIOMET, INC.
Product Code
LZO
UDI-DI
00880304514393
PMA / PMN Number
K200196
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE FOLLOWING SECTIONS WERE UPDATED/CORRECTED UPDATED: B4 G3 G6 H2 H6 H11. THE DEVICE EVALUATION FOUND THE PRODUCT TO BE CONFORMING AND STERILITY NOT COMPROMISED. FURTHER, THE EVENT DID NOT CONTRIBUTE TO A SERIOUS INJURY; THEREFORE, THIS WOULD NOT BE CONSIDERED A REPORTABLE EVENT. VISUAL EVALUATION OF THE RETURNED PRODUCT/PROVIDED PHOTOS CONFIRMED THAT STERILITY WAS NOT BREACHED. REVIEW OF THE DEVICE HISTORY RECORDS IDENTIFIED NO DEVIATIONS OR ANOMALIES DURING MANUFACTURING RELATED TO THE REPORTED EVENT. THE REPORTED EVENT DID NOT OCCUR IN AN OPERATING ROOM OR AS PART OF A MEDICAL PROCEDURE; MEDICAL RECORDS ARE NOT AVAILABLE FOR REVIEW. THE CONDITION OF THE DEVICE WHEN IT LEFT ZIMMER BIOMET IS CONSIDERED CONFORMING TO SPECIFICATION. THE ROOT CAUSE OF THE REPORTED EVENT CAN BE ATTRIBUTED TO TRANSIT DAMAGE AND A PACKAGING DESIGN ISSUE. THE REPORTED EVENT HAS BEEN CONFIRMED BY EVALUATION OF THE RETURNED PRODUCT AND PROVIDED PHOTOS. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Additional Manufacturer Narrative · 0

(B)(4). D10: 51-104140 ITEM NAME TPRLC 133 T1 PPS HO 14X148MM LOT # 7142515 51-104140 ITEM NAME TPRLC 133 T1 PPS HO 14X148MM LOT # 7142515 51-103160 ITEM NAME TPRLC 133 T1 PPS SO 16X152MM LOT # 6170598 G2: FOREIGN: JAPAN IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 0

THERE IS NO UPDATE TO THE PRIOR EVENT DESCRIPTION PROVIDED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE STERILE PACKAGING WAS FOUND DAMAGED. THERE IS NO ADDITIONAL INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
681910 TPRLC 133 MP TYPE1 PPS SO 16.0 PROSTHETIC, HIP LZO ZIMMER BIOMET, INC. 7208572 00880304514393

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown