FDA Adverse Event Injury Summary report: N

EXCELSIUS GPS

MDR report key: 21264882 · Received January 29, 2025

Report

Report Number
3004142400-2024-00420
Event Type
Injury
Date Received
January 29, 2025
Date of Event
March 28, 2023
Report Date
January 29, 2025
Manufacturer
GLOBUS MEDICAL, INC.
Product Code
OLO
UDI-DI
00889095481297
PMA / PMN Number
K171651
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS IS BEING REPORTED FOR A PROBLEM FOUND EARLIER. INVESTIGATION REVEALED THAT THERE WAS NO SYSTEM MALFUNCTION. THIS ERROR GETS RAISED WHEN ONE OR MORE AXES ARE NOT IN STANDSTILL STATE WHEN TRYING TO START MOVEMENT. MOTION WAS LOST DURING THE CASE BECAUSE WHEN ANY AXIS IS IN AN ERROR STATE, ALL AXES ARE PREVENTED FROM MOVING. IMPROVEMENTS TO GMAS SOFTWARE WERE MADE TO ADDRESS THIS ISSUE IN THE SOFTWARE RELEASE FOLLOWING THIS INCIDENT. IN PARTICULAR, THE CODE THAT POWERS AXE OFF WAS MADE TO BE MORE ROBUST, ENSURING AXES WOULD UNDERGO THE CORRECT STATE TRANSITIONS. AS NOTED IN THE ATTACHED RISK RECONCILIATION FORM, THE OBSERVED RISK LEVEL IS LOW AND IS EQUAL TO THE ANTICIPATED RISK LEVEL . THE OVERALL RISK OF THE SYSTEM HAS BEEN MAINTAINED AS "LOW" AN NO FURTHER INVESTIGATION IS REQUIRED. THE CAUSE OF THE REPORTED ISSUE CAN BE TRACED TO USER TECHNIQUE.

Description of Event or Problem · 0

L4-5 INTRA-OP CASE. REGISTRATION NORMAL/NO ERRORS/NO SHIFTS. SCREW PLACEMENT: L4-R PLACED, SELECTED L5-R, PRESSED FOOT PEDAL, EE CAME OFF L4-R LIKE NORMAL. MOVED EE OUT OF WAY WITH RING. ADJUSTED L5-R PLAN, BACK TO NAVIGATE, PRESSED FOOT PEDAL, ARM FLOATED TO THE SIDE. THEN ARM LOCKED UP/WOULD NOT MOVE VIA BUTTONS, FOOT PEDAL, OR RING. ARM WOULD NOT MOVE, RE-HOME, OR RE-CALIBRATE LOADCELL AFTER MULTIPLE HARD SHUTDOWNS AND SOFTWARE RESETS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2047328 EXCELSIUS GPS SYSTEM, EXCELSIUS GPS, 120V OLO GLOBUS MEDICAL, INC. 6143.1001 GPS-0159 00889095481297

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other