FDA Adverse Event Malfunction Summary report: N

DAVINCI XI

MDR report key: 21260817 · Received January 28, 2025

Report

Report Number
2955842-2025-00937
Event Type
Malfunction
Date Received
January 28, 2025
Date of Event
January 3, 2025
Report Date
January 3, 2025
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874110898
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN INVESTIGATION WAS COMPLETED TO DETERMINE THE CAUSE OF THIS REPORTED EVENT. AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. THE FSE REPLACED THE INTEGRATED ELECTROSURGICAL UNIT (IESU) TO RESOLVE THE ISSUE. THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. ISI DID RECEIVE A DA VINCI PRODUCT INVOLVED WITH THIS COMPLAINT TO PERFORM FAILURE ANALYSIS. THE IESU ENERGIZED AND CAUTERIZED AND ALL PORTS RECOGNIZED INSTRUMENTS ON SYSTEM. THE M-11, M-18, AND C-06 ERRORS WERE IN THE LOGS. THE ROOT CAUSE IS ATTRIBUTED TO A DEFECTIVE COMPONENT.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED RADICAL PROSTATECTOMY SURGICAL PROCEDURE THAT THE INTEGRATED ELECTROSURGICAL UNIT (IESU) FAULTED WHILE TRYING TO USE MONOPOLAR ENERGY. THE SYSTEM LOGS CONFIRMED THAT IESU ERRORS C-06, M-18, AND M-11 OCCURRED. THE INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SUPPORT ENGINEER (TSE) INQUIRED IF ANY ENVIRONMENTAL SOURCES OF ELECTROMAGNETIC INTERFERENCE WERE IN CLOSE PROXIMITY TO THE IESU; THE CUSTOMER CONFIRMED THERE WAS NONE. THE TSE RECOMMENDED DISCONNECTING THE IESU FOOTSWITCH PEDALS, BUT THE CUSTOMER DECLINED AS THE PROCEDURE WAS ONGOING. THERE WERE NO REPORTS OF PATIENT INJURY. PROCEDURE WAS COMPLETED. ISI PERFORMED FOLLOW-UP AND OBTAINED PATIENT RELATED INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1995863 DAVINCI XI VISION SIDE SYSTEM NAY INTUITIVE SURGICAL, INC 381121-21 N/A 00886874110898

Patients

Seq Age Sex Outcome Treatment
1 73 YR Male DA VINCI INSTRUMENTS AND ACCESSORIES.