FDA Adverse Event Malfunction Summary report: N

ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT KIT

MDR report key: 21257068 · Received January 28, 2025

Report

Report Number
3005094123-2025-00041
Event Type
Malfunction
Date Received
January 28, 2025
Date of Event
January 20, 2025
Report Date
February 20, 2025
Manufacturer
ABBOTT IRELAND DIAGNOSTICS DIVISION
Product Code
MMI
UDI-DI
00380740186029
PMA / PMN Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AFTER FURTHER EVALUATION, THE SUSPECT MEDICAL DEVICE WAS CHANGED FROM ALINITY I STAT HIGH SENSITIVE TROPONIN-I, 08P13, LONGFORD MANUFACTURING SITE TO ALINITY I PROCESSING MODULE, 03R65, IRVING IA/CC MANUFACTURING SITE. MDR NUMBER 3016438761-2025-00113 HAS BEEN SUBMITTED AND ALL FURTHER INFORMATION WILL BE DOCUMENTED UNDER THAT MDR NUMBER.

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 8P13 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 4Z21, WITH 510K/PMA/BLA NUMBER K202525.

Description of Event or Problem · 0

THE CUSTOMER REPORTED A FALSE POSITIVE ALINITY STAT HIGH SENSITIVE TROPONIN-I RESULT FOR ONE PATIENT GENERATED ON THE ALINITY I PROCESSING MODULE. THE FOLLOWING DATA WAS PROVIDED: SID (B)(6) INITIAL RESULT = 46 NG/L REPEAT RESULT = 3.8 NG/L REPEAT RESULT ON ANOTHER MODULE = 3.8 NG/L. NO CUSTOMER CUTOFF WAS PROVIDED FOR ALINITY I STAT HIGH SENSITIVE TROPONIN-I, THEREFORE USING THE PACKAGE INSERT OVERALL CUT OFF OF 26.2 NG/L. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED A FALSE POSITIVE ALINITY STAT HIGH SENSITIVE TROPONIN-I RESULT FOR ONE PATIENT GENERATED ON THE ALINITY I PROCESSING MODULE. THE FOLLOWING DATA WAS PROVIDED: SID (B)(6). INITIAL RESULT = 46 NG/L, REPEAT RESULT = 3.8 NG/L, REPEAT RESULT ON ANOTHER MODULE = 3.8 NG/L. NO CUSTOMER CUTOFF WAS PROVIDED FOR ALINITY I STAT HIGH SENSITIVE TROPONIN-I, THEREFORE USING THE PACKAGE INSERT OVERALL CUT OFF OF 26.2 NG/L. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2377255 ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT KIT IMMUNOASSAY METHOD, TROPONIN SUBUNIT MMI ABBOTT IRELAND DIAGNOSTICS DIVISION 68456UD00 00380740186029

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown ALNTY I PROCESSING MODU, 03R65-01, (B)(6)| ALNTY I PROCESSING MODU, 03R65-01, (B)(6)