ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT KIT
Report
- Report Number
- 3005094123-2025-00041
- Event Type
- Malfunction
- Date Received
- January 28, 2025
- Date of Event
- January 20, 2025
- Report Date
- February 20, 2025
- Manufacturer
- ABBOTT IRELAND DIAGNOSTICS DIVISION
- Product Code
- MMI
- UDI-DI
- 00380740186029
- PMA / PMN Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
AFTER FURTHER EVALUATION, THE SUSPECT MEDICAL DEVICE WAS CHANGED FROM ALINITY I STAT HIGH SENSITIVE TROPONIN-I, 08P13, LONGFORD MANUFACTURING SITE TO ALINITY I PROCESSING MODULE, 03R65, IRVING IA/CC MANUFACTURING SITE. MDR NUMBER 3016438761-2025-00113 HAS BEEN SUBMITTED AND ALL FURTHER INFORMATION WILL BE DOCUMENTED UNDER THAT MDR NUMBER.
AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 8P13 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 4Z21, WITH 510K/PMA/BLA NUMBER K202525.
THE CUSTOMER REPORTED A FALSE POSITIVE ALINITY STAT HIGH SENSITIVE TROPONIN-I RESULT FOR ONE PATIENT GENERATED ON THE ALINITY I PROCESSING MODULE. THE FOLLOWING DATA WAS PROVIDED: SID (B)(6) INITIAL RESULT = 46 NG/L REPEAT RESULT = 3.8 NG/L REPEAT RESULT ON ANOTHER MODULE = 3.8 NG/L. NO CUSTOMER CUTOFF WAS PROVIDED FOR ALINITY I STAT HIGH SENSITIVE TROPONIN-I, THEREFORE USING THE PACKAGE INSERT OVERALL CUT OFF OF 26.2 NG/L. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
THE CUSTOMER REPORTED A FALSE POSITIVE ALINITY STAT HIGH SENSITIVE TROPONIN-I RESULT FOR ONE PATIENT GENERATED ON THE ALINITY I PROCESSING MODULE. THE FOLLOWING DATA WAS PROVIDED: SID (B)(6). INITIAL RESULT = 46 NG/L, REPEAT RESULT = 3.8 NG/L, REPEAT RESULT ON ANOTHER MODULE = 3.8 NG/L. NO CUSTOMER CUTOFF WAS PROVIDED FOR ALINITY I STAT HIGH SENSITIVE TROPONIN-I, THEREFORE USING THE PACKAGE INSERT OVERALL CUT OFF OF 26.2 NG/L. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2377255 | ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT KIT | IMMUNOASSAY METHOD, TROPONIN SUBUNIT | MMI | ABBOTT IRELAND DIAGNOSTICS DIVISION | 68456UD00 | 00380740186029 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | ALNTY I PROCESSING MODU, 03R65-01, (B)(6)| ALNTY I PROCESSING MODU, 03R65-01, (B)(6) |