FDA Adverse Event Malfunction Summary report: N

IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 8MM

MDR report key: 21250732 · Received January 28, 2025

Report

Report Number
0002023141-2025-00230
Event Type
Malfunction
Date Received
January 28, 2025
Report Date
April 2, 2025
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024020085
PMA / PMN Number
K061410
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). B3: DATE OF EVENT UNKNOWN / NOT PROVIDED. D9: DEVICE AVAILABILITY UNKNOWN / NOT PROVIDED. G4: PREMARKET IDENTIFICATION K011028/K013227.

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). THE FOLLOWING FIELDS HAVE BEEN UPDATED: B4: DATE OF THIS REPORT, B5: DESCRIBE EVENT OR PROBLEM, D9: DEVICE AVAILABILITY, G3: DATE RECEIVED BY MANUFACTURER, G6: TYPE OF REPORT, H1: TYPE OF REPORTABLE EVENT, H2: FOLLOW UP TYPE, H3: DEVICE EVALUATED BY MANUFACTURER, H6: ADVERSE EVENT PROBLEM, H10: ADDITIONAL NARRATIVE. ZIMVIE RECEIVED ONE (1) TSVWB8, (IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 8MM) FOR EVALUATION. VISUAL EVALUATION WAS PERFORMED TO THE IMPLANT AND THERE WAS DAMAGE LIKELY FROM THE REMOVAL PROCESS. THE MOUNTS HEX WAS FRACTURED AND STUCK INSIDE THE IMPLANT. DEVICE HISTORY RECORD (DHR) REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1251406. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1251406 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: DENTAL: FUNCTIONAL: DOES NOT DISENGAGE/RELEASE AND DENTAL: FUNCTIONAL: FRACTURE: MOUNT BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW, THE MOST LIKELY ROOT CAUSE DETERMINED FROM THE INVESTIGATION WAS THE CLINICIAN DIDN¿T FOLLOW RECOMMENDED PROTOCOL FOR IMPLANT PLACEMENT AND FINAL SEATING. THEREFORE, BASED ON THE AVAILABLE INFORMATION, A DEVICE MALFUNCTION DID OCCUR. THE MOUNT WAS FRACTURED AT THE HEX. THE REPORTED EVENT WAS CONFIRMED. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCTS WERE NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED. AT THIS TIME, THE COMPLAINT INVESTIGATION HAS BEEN COMPLETED AND THE RECORD WILL BE CLOSED. IF ADDITIONAL INFORMATION IS RECEIVED, THE RECORD WILL BE RE-OPENED FOR FURTHER EVALUATION.

Description of Event or Problem · 0

WHEN THE IMPLANT WAS PLACED, THE MOUNT REMAINED ATTACHED TO THE IMPLANT. AFTER SEVERAL ATTEMPTS, PART OF THE MOUNT FRACTURED INSIDE THE IMPLANT. IT WAS THEREFORE REMOVED AND REPLACED BY A LARGER IMPLANT.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1549237 IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 8MM DENTAL IMPLANT DZE ZIMMER DENTAL 1251406 00889024020085

Patients

Seq Age Sex Outcome Treatment
1 26 YR Female