FDA Adverse Event Malfunction Summary report: N

AGILIA VP MC WIFI

MDR report key: 21246439 · Received January 27, 2025

Report

Report Number
3004548776-2025-00034
Event Type
Malfunction
Date Received
January 27, 2025
Date of Event
January 20, 2025
Report Date
November 5, 2025
Manufacturer
FRESENIUS VIAL S.A.S.
Product Code
FRN
Product Problem
Yes
Report Source
Distributor report
Reporter Location
SW
Reporter Occupation
OTHER
Health Professional
N

Narratives

Description of Event or Problem · 0

NO ISSUE COULD BE FOUND DURING DEVICE HISTORY RECORD REVIEW. DEVICE LOG HISTORY REGARDING THE SAME REPORTED EVENT WAS REVIEWED. SEVERAL AIR ALARMS WERE IDENTIFIED BUT ALTHOUGH THIS COULD CONFIRM THE REPORTED EVENT (AIR IN LINE), THOSE INFORMATION ARE INSUFFICIENT TO CONFIRM A QUALITY ISSUE AS ALARMS ARE DESIGNED TO INFORM THE USER THAT THE INFUSION NEEDS TO BE ATTENDED. A REPORTED DEVICE AND 5 DEFECTIVE AIR SENSORS WERE RECEIVED IN BRÉZINS FOR INVESTIGATION. A LENGTHY INVESTIGATION INTO EVENT 2043114 WAS CARRIED OUT BY THE R&D TEAM. IT HAS BEEN SHOWN THAT THE AIR DETECTION LEVELS OF THE LAST SOFT (SW. 4.2 AND 4.3) HAVE BEEN LOWERED IN ORDER TO BETTER DETECT SMALL BUBBLES (<10 L) WHICH WORKS VERY WELL WITH NEW VERSIONS OF AIR SENSORS. THE ISSUE IS THAT WITH SLIGHTLY OLDER AIR SENSORS, THE DETECTION THRESHOLD IS TOO CLOSE TO THE LIMIT, TOO SENSITIVE. THEREFORE, SEVERAL ACTIONS ARE BEING CARRIED OUT. AN FSN HAS BEEN SENT, FRESENIUS KABI RECOMMENDS TO ALL CUSTOMERS: ·WHEN RELOCATING PUMP STATIONS, ALWAYS MOVE THE ROLLING STAND ITSELF, RATHER THAN PULLING ON THE TUBE SETS. ·WHEN INSTALLING TUBE SETS, TO USE THE HOOK ON THE LEFT SIDE OF THE PUMP HOUSING TO SECURE THE TUBE SET AND KEEP IT IN PLACE THROUGHOUT THE INFUSION. A SOFTWARE CORRECTION TO PREVENT FROM THESE FALSE AIR ALARMS IS IN PREPARATION. A HARDWARE INVESTIGATION IS ALSO UNDERWAY ON THE AIR SENSORS IN THE FIELD VIA AN INTERNAL CAPA. WE REMIND YOU THAT BEFORE UPGRADING, IT IS IMPORTANT TO FOLLOW THE TSB0455. THIS COMPLAINT IS CONSIDERED AS VALID, AND THE ROOT CAUSE IS RELATED TO A SOFTWARE ISSUE. TO OUR KNOWLEDGE THIS COMPLAINT IS NOT BEING RELATED TO PATIENT SAFETY. THIS COMPLAINT IS CONSIDERED AS: VALID. THE TREND IS: ABNORMAL.

Description of Event or Problem · 0

THE FOLLOWING HAS BEEN REPORTED: AN INCREASE OF AIR ALARMS HAS BEEN NOTICED BY SEVERAL INDEPENDENT DEPARTMENTS AFTER UPGRADING THE VP PUMP FROM SW 4.1 TO 4.2B. THE ALARM RATE IS A LOT HIGHER THAT BEFORE AND NO AIR IN THE LINES ARE FOUND AFTER RECEIVING THE ALARM ACCORDING TO THE CLIENTS. REPORTING AS A CONSERVATIVE MEASURE. NO ADVERSE EFFECTS WERE REPORTED. MORE INFORMATION IS NEEDED TO COMPLETE THE INVESTIGATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
630378 AGILIA VP MC WIFI INFUSION PUMP SYSTEM FRN FRESENIUS VIAL S.A.S. Z019700

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown