FDA Adverse Event Malfunction Summary report: N

NC QUANTUM APEX PTCA DILATATION CATHETER

MDR report key: 2123994 · Received June 14, 2011

Report

Report Number
2134265-2011-02259
Event Type
Malfunction
Date Received
June 14, 2011
Date of Event
May 22, 2011
Report Date
May 22, 2011
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LOX
PMA / PMN Number
P860019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4)DEVICE EVALUATED BY MFR: THE DEVICE HAS NOT BEEN RECEIVED FOR ANALYSIS. UPON RECEIPT AND COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED.(B)(4)

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR: EXAMINATION OF THE RETURNED DEVICE REVEALED BLOOD AND CONTRAST IN THE LUMEN AND BALLOON. THERE WAS A LONGITUDINAL TEAR BEGINNING AT THE PROXIMAL BALLOON WAIST EXTENDING 12MM DISTALLY. THE WIRE LUMEN SHAFT WAS BUCKLED 25MM FROM THE PROXIMAL BALLOON WAIST. THE DISTAL END OF THE BALLOON WAS BUNCHED UP. INSPECTION OF THE BALLOON MATERIAL AT THE EDGES OF THE TEAR PRESENTED NO INDICATION OF AN INITIATION POINT OR ANY MATERIAL OR MANUFACTURING DEFICIENCIES THAT COULD HAVE CONTRIBUTED TO THE FORMATION OF THE TEAR. INSPECTION OF THE REMAINDER OF THE DEVICE PRESENTED NO OTHER DAMAGE OR IRREGULARITIES. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY TREATMENT PROCEDURE, A BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY CALCIFIED AND MODERATELY TORTUOUS PROXIMAL LEFT ANTERIOR DESCENDING ARTERY. THE 2.5X15MM NC QUANTUM APEX BALLOON CATHETER WAS ADVANCED TO THE LESION. DURING THE FIRST INFLATION, THE BALLOON RUPTURED AT 18ATMS. THE DEVICE WAS REMOVED INTACT AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT'S STATUS IS GOOD.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY TREATMENT PROCEDURE, A BALLOON RUPTURE OCCURRED.THE 90% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY CALCIFIED AND MODERATELY TORTUOUS PROXIMAL LEFT ANTERIOR DESCENDING ARTERY. THE 2.5X15MM NC QUANTUM APEX BALLOON CATHETER WAS ADVANCED TO THE LESION. DURING THE FIRST INFLATION, THE BALLOON RUPTURED AT 18ATMS. THE DEVICE WAS REMOVED INTACT AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT'S STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NC QUANTUM APEX PTCA DILATATION CATHETER CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - MAPLE GROVE H7493912415250 13878849

Patients

Seq Age Sex Outcome Treatment
1