FDA Adverse Event Malfunction Summary report: N

ATTAIN ABILITY PLUS

MDR report key: 2123991 · Received June 14, 2011

Report

Report Number
2649622-2011-08213
Event Type
Malfunction
Date Received
June 14, 2011
Date of Event
March 3, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
OJX
PMA / PMN Number
P080006/S002
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT ATTEMPT, WHEN REMOVING THE STYLET, THE INSULATION ON THE IS-1 PIN APPEARED TO BE DAMAGED. THE LEAD WAS NOT IMPLANTED AND ANOTHER LEAD WAS USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN ABILITY PLUS IMPLANTABLE PACING LEAD OJX MEDTRONIC PUERTO RICO, INC. 4296 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Other