FDA Adverse Event
Injury
Summary report: N
SECURA DR
MDR report key: 2123987
·
Received June 14, 2011
Report
- Report Number
- 6000144-2011-02310
- Event Type
- Injury
- Date Received
- June 14, 2011
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- LWS
- PMA / PMN Number
- P980016/S114
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PATIENT ALERT SOUNDED, AND THERE WAS A POSSIBLE CONNECTOR ISSUE CAUSING NOISE. MANIPULATION OF THE DEVICE DID NOT REPRODUCE THE NOISE. THE DEVICE WAS REPROGRAMMED AND IS STILL IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SECURA DR | IMPLANTABLE PACEMAKER/CARDIO/DEFIB | LWS | MEDTRONIC MED REL, INC. | D224DRG | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 20 YR | Other| R | (B)(4) COMPETITOR IMPLANTABLE PACING LEAD| (B)(4) TISSUE VALVE| (B)(4) COMPETITOR IMPLANTABLE TACHY LEAD |