CAPSUREFIX NOVUS
MDR report key: 2123924
·
Received June 14, 2011
Report
- Report Number
- 2649622-2011-08035
- Event Type
- Injury
- Date Received
- June 14, 2011
- Date of Event
- March 3, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P930039/S009
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD IN/ON THE HELIX MECHANISM. (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD IN/ON THE HELIX MECHANISM.
Description of Event or Problem · 1
IT WAS REPORTED THAT BOTH THE ATRIAL AND VENTRICULAR LEADS HAD DISLODGED. IT WAS ALSO REPORTED THAT THE RIGHT VENTRICULAR LEAD HAD UNACCEPTABLE THRESHOLDS. THE LEADS WERE EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5076 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 26 YR | Hospitalization| R | VEDR01 IMPLANTABLE PACEMAKER/CARDIO/DEFIB |