FDA Adverse Event Malfunction Summary report: N

GYNECARE X-TRACT MORCELLATOR

MDR report key: 2123903 · Received June 14, 2011

Report

Report Number
2210968-2011-00770
Event Type
Malfunction
Date Received
June 14, 2011
Date of Event
May 24, 2011
Report Date
May 25, 2011
Manufacturer
ETHICON
Product Code
HET
PMA / PMN Number
K993801
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CONCLUSION: THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY AND FUNCTIONALLY EVALUATED. THE REPORTED SYMPTOM WAS DUPLICATED. THE CAUSE WAS DUE TO DAMAGED HARDWARE. IN ADDITION, A REVIEW OF THE DEVICE MANUFACTURING RECORDS WAS CONDUCTED AND THE DEVICE MET ALL FINISHED GOODS RELEASE CRITERIA.

Additional Manufacturer Narrative · 1

(B)(4). UNABLE TO ACTIVATE DISPOSABLE. CONCLUSION: THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED HAS BEEN RECEIVED, HOWEVER, THE PRODUCT EVALUATION IS NOT YET COMPLETE. ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT A GYNECOLOGICAL PROCEDURE ON (B)(6) 2011. DURING THE PROCEDURE, WHEN THE TRIGGER ON THE DEVICE WAS PRESSED, THE BLADE WOULD NOT SPIN. A SECOND MOTOR DRIVE UNIT WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GYNECARE X-TRACT MORCELLATOR LAPAROSCOPIC MORCELLATOR HET ETHICON NA NA

Patients

Seq Age Sex Outcome Treatment
1