FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEM

MDR report key: 2123900 · Received June 14, 2011

Report

Report Number
2050012-2011-02287
Event Type
Malfunction
Date Received
June 14, 2011
Date of Event
May 17, 2011
Report Date
May 17, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2011. THE FSE OBSERVED REAGENT PROBE VACUUM VALVE CORRODED, AND REPLACED THE VACUUM VALVE. HARDWARE IS THE ROOT CAUSE FOR THIS EVENT.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) AND REPORTED THAT FLUID WAS FOUND ON THE REAGENT CARTRIDGE AND THE REAGENT PROBE B IS LEAKING. CUSTOMER CHANGED THE REAGENT PROBE B OUT AND THE PROBE WAS STILL LEAKING. NO INJURY WAS REPORTED ON THIS ISSUE.