FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2123898 · Received June 14, 2011

Report

Report Number
2649622-2011-08026
Event Type
Injury
Date Received
June 14, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE PROXIMAL SEGMENT OF THE DEVICE WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. IT WAS ALSO NOTED THAT THE OUTER INSULATION HAD A COSMETIC DEPRESSION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT HEARD AN ALERT TONE FOR HIGH IMPEDANCE AND THAT THE SUPERIOR VENA CAVA HIGH VOLTAGE COIL OF THE LEAD WAS APPARENTLY FRACTURED. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 68 YR Hospitalization| R 4193 IMPLANTABLE PACING LEAD| 7304 IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD