FDA Adverse Event Other Summary report: N

DA VINCI S SURGICAL SYSTEM

MDR report key: 2123830 · Received June 14, 2011

Report

Report Number
2955842-2011-00176
Event Type
Other
Date Received
June 14, 2011
Date of Event
May 16, 2011
Report Date
May 16, 2011
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K081207
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION CONDUCTED BY ISI FIELD SERVICE ENGINEERING FOUND THE POWER ISSUE EXPERIENCED BY THE CUSTOMER WAS ASSOCIATED WITH THE SURGEON SIDE CONSOLE PRIMARY POWER SUPPLY (PPS2). THE PPS2 IS A 230V POWER SUPPLY WITH BATTERY BACKUP. THE SYSTEM WAS REPAIRED BY REPLACING THE PPS2 AND THE POWER SUPPLY FAN AND THE BATTERY PACK MODULE AS A PRECAUTION. AS OF (B)(4) 2011, THERE HAVE BEEN NO REPORTED RECURRENCES OF THE ISSUE AT THIS HOSPITAL.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI S LOW ANTERIOR RESECTION PROCEDURE, THE SURGEON SIDE CONSOLE (SSC) EXPERIENCED A POWER LOSS. WITH THE ASSISTANCE OF AN ISI TECHNICAL SUPPORT ENGINEER, THE SITE RESTARTED THE SYSTEM, HOWEVER, AFTER HOMING THE SSC CONTINUED TO RUN OFF OF THE BATTERY BACK-UP SYSTEM. THE SITE TRIED MULTIPLE OUTLETS IN THE OPERATING ROOM, HOWEVER, THE SSC CONTINUED TO RUN FROM THE BATTERY BACK-UP SYSTEM. THE SURGEON DECIDED TO CONVERT TO TRADITIONAL OPEN SURGICAL TECHNIQUES TO FINISH THE PLANNED PROCEDURE. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DA VINCI S SURGICAL SYSTEM ENDOSCOPIC INSTRUMENT CONTROL SYSTEM NAY INTUITIVE SURGICAL,INC. IS2000 A5.1P8

Patients

Seq Age Sex Outcome Treatment
1 Other DA VINCI S SYSTEM INSTRUMENTS AND ACCESSORIES