DA VINCI S SURGICAL SYSTEM
Report
- Report Number
- 2955842-2011-00176
- Event Type
- Other
- Date Received
- June 14, 2011
- Date of Event
- May 16, 2011
- Report Date
- May 16, 2011
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K081207
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE INVESTIGATION CONDUCTED BY ISI FIELD SERVICE ENGINEERING FOUND THE POWER ISSUE EXPERIENCED BY THE CUSTOMER WAS ASSOCIATED WITH THE SURGEON SIDE CONSOLE PRIMARY POWER SUPPLY (PPS2). THE PPS2 IS A 230V POWER SUPPLY WITH BATTERY BACKUP. THE SYSTEM WAS REPAIRED BY REPLACING THE PPS2 AND THE POWER SUPPLY FAN AND THE BATTERY PACK MODULE AS A PRECAUTION. AS OF (B)(4) 2011, THERE HAVE BEEN NO REPORTED RECURRENCES OF THE ISSUE AT THIS HOSPITAL.
IT WAS REPORTED THAT DURING A DA VINCI S LOW ANTERIOR RESECTION PROCEDURE, THE SURGEON SIDE CONSOLE (SSC) EXPERIENCED A POWER LOSS. WITH THE ASSISTANCE OF AN ISI TECHNICAL SUPPORT ENGINEER, THE SITE RESTARTED THE SYSTEM, HOWEVER, AFTER HOMING THE SSC CONTINUED TO RUN OFF OF THE BATTERY BACK-UP SYSTEM. THE SITE TRIED MULTIPLE OUTLETS IN THE OPERATING ROOM, HOWEVER, THE SSC CONTINUED TO RUN FROM THE BATTERY BACK-UP SYSTEM. THE SURGEON DECIDED TO CONVERT TO TRADITIONAL OPEN SURGICAL TECHNIQUES TO FINISH THE PLANNED PROCEDURE. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DA VINCI S SURGICAL SYSTEM | ENDOSCOPIC INSTRUMENT CONTROL SYSTEM | NAY | INTUITIVE SURGICAL,INC. | IS2000 A5.1P8 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other | DA VINCI S SYSTEM INSTRUMENTS AND ACCESSORIES |