CD HORIZON SOLERA 5.5/6.0
Report
- Report Number
- 1030489-2025-00383
- Event Type
- Malfunction
- Date Received
- January 26, 2025
- Date of Event
- June 5, 2023
- Report Date
- January 26, 2025
- Manufacturer
- MEDTRONIC SOFAMOR DANEK USA, INC
- Product Code
- HXX
- UDI-DI
- 00643169043640
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- 003
Narratives
H3: PRODUCT ANALYSIS: PART # (B)(4); LOT # K23B1472 VISUAL INSPECTION CONFIRMED THE ENTIRE TORX TIP OF INSTRUMENT HAS BEEN SHEARED OFF AND THE ATTACHMENT PIN TO THE INTERNAL BUSHING HAS BEEN BROKEN OFF, CONSISTENT WITH INTERFACE DURING USAGE. INSPECTION OF THE SHAFT DIAMETER AND MATERIAL HARDNESS CONFIRMED CONFORMANCE TO PRINT SPECIFICATION. OPTICAL EXAMINATION OF THE FRACTURE SURFACE REVEALED A FAIRLY FLAT FRACTURE SURFACE AND CIRCULAR MATERIAL FLOW, CONSISTENT WITH TORSIONAL OVERLOAD. THIS TYPE OF DAMAGE IS CONSISTENT WITH TORSIONAL OVERLOAD. THIS REGULATORY REPORT IS BEING SUBMITTED DUE TO RETROSPECTIVE REVIEW THROUGH CAPA 624392. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
INFORMATION WAS RECEIVED VIA MANUFACTURER REPRESENTATIVE REGARDING A DRIVER USED IN SPINAL THERAPY. IT WAS REPORTED THAT THE REP NOTICED DAMAGED TIP AFTER THE INSTRUMENT CAME OUT OF THE WASH. NO PATIENT WAS INVOLVED IN THE EVENT. NO FURTHER COMPLICATIONS WERE REPORTED/ANTICIPATED. IT WAS REPORTED THAT IT WAS BREAKAGE ASSOCIATED WITH UNDESIRED DAMAGE OR BREAKAGE OF THE TIP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1606691 | CD HORIZON SOLERA 5.5/6.0 | SCREWDRIVER | HXX | MEDTRONIC SOFAMOR DANEK USA, INC | 5584111 | K23B1472 | 00643169043640 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |