LIFEVEST WCD 3100 SYSTEM
Report
- Report Number
- 3008642652-2011-00018
- Event Type
- Malfunction
- Date Received
- June 10, 2011
- Date of Event
- April 18, 2011
- Report Date
- June 8, 2011
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BATTERY PACK (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (BATTERY WON'T HOLD CHARGE) HAS BEEN CONFIRMED. THE CAUSE OF THE DISCHARGED BATTERY PACK WAS DEFECTIVE CELLS WITHIN THE BATTERY PACK. ONE OR MORE OF THE CELLS HAD RESIDUE/CORROSION ON THEM. THE SOURCE OF THE RESIDUE/CORROSION HAS NOT BEEN POSITIVELY IDENTIFIED BUT IS LIKELY ELECTROLYTE FROM A LEAKING CELL. THE ROOT CAUSE OF THE LEAKING CELL WAS NOT POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY PACK. THE PATIENT RECEIVED A REPLACEMENT BATTERY PACK.
ZOLL CUSTOMER SUPPORT CONTACTED A (B)(6) MALE PATIENT IN REGARDS TO A DOWNLOAD WHICH REVEALED EXCESSIVE BATTERY CHARGER FAULTS. THE PATIENT WAS ISSUED A REPLACEMENT BATTERY PACK.