CERENOVUS ENTERPRISE
Report
- Report Number
- 3008114965-2025-00054
- Event Type
- Malfunction
- Date Received
- January 24, 2025
- Date of Event
- January 9, 2025
- Report Date
- February 11, 2025
- Manufacturer
- MEDOS INTERNATIONAL SARL
- Product Code
- NJE
- UDI-DI
- 10886704043980
- PMA / PMN Number
- H60001
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
MANUFACTURER¿S REF. NO: (B)(4). THE PURPOSE OF THIS MDR SUBMISSION IS TO REPORT THE INVESTIGATIONAL FINDING OF THE RETURNED DEVICE. THE COMPLAINT PRODUCT WAS RETURNED AND RECEIVED FOR EVALUATION AND ANALYSIS. THE INVESTIGATION IS DOCUMENTED BELOW. INVESTIGATION SUMMARY: A NON-STERILE 4.5MM X 22MM ENTERPRISE® VASCULAR RECONSTRUCTION DEVICE WAS RECEIVED CONTAINED IN THE DECONTAMINATION POUCH. VISUAL INSPECTION WAS PERFORMED. AS DESCRIBED IN THE COMPLAINT EVENT DESCRIPTION, THE STENT COMPONENT WAS DETACHED FROM THE DELIVERY SYSTEM. THE DISTAL END OF THE DELIVER WIRE WAS BROKEN NEXT TO THE RETRACTING BUMP. NO OTHER DAMAGES WERE OBSERVED. MICROSCOPIC INSPECTION WAS PERFORMED ON THE STENT COMPONENT. IT WAS OBSERVED TO BE IN GOOD CONDITION; THERE WAS NO STRUCTURAL DAMAGE (I.E., NO BROKEN STRUTS, NO KINKS); IT WAS ALSO NOTED TO BE FULLY EXPANDED WITH BOTH ENDS COMPLETELY FLARED. THE DISTAL COIL OF THE DELIVERY WIRE WAS FOUND STRETCHED. THE BROKEN CONDITION OF THE DELIVERY WIRE SUGGEST THAT THE DELIVERY WIRE WAS PUSHED OR RETRACTED AGAINST RESISTANCE SUFFICIENTLY TO CAUSE THE DELIVERY WIRE TO BREAK AND RELEASE THE STENT INSIDE THE Y-CONNECTOR. THEREFORE, THE REPORTED ISSUE DOCUMENTED IN THE COMPLAINT CAN BE CONFIRMED. THERE IS NO INDICATION THAT THE ISSUE REPORTED IN THE COMPLAINT IS A RESULT OF A DEFECT INHERENTLY RELATED TO THE DEVICE. LAKE REGION MEDICAL PERFORMED A REVIEW OF THE DEVICE HISTORY RECORDS RELATIVE TO THE MANUFACTURING, INSPECTION, AND PACKAGING OF THE LOT 9080733. THE HISTORY RECORD INDICATES THIS PRODUCT WAS FINAL INSPECTION TESTED AT LAKE REGION MEDICAL AND WAS DETERMINED TO BE ACCEPTABLE. AS PART OF JOHNSON & JOHNSON MEDTECH QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. IT SHOULD BE NOTED THAT PRODUCT FAILURE COULD BE CAUSED BY MULTIPLE FACTORS. THE INSTRUCTIONS FOR USE (IFU) DOES CONTAIN THE FOLLOWING RECOMMENDATIONS: IF RESISTANCE IS MET DURING MANIPULATION, DETERMINE THE CAUSE OF RESISTANCE BEFORE PROCEEDING. DO NOT APPLY UNDUE FORCE IF RESISTANCE IS ENCOUNTERED AT ANY POINT DURING STENT MANIPULATION. WITHDRAW THE UNIT AND ADVANCE A NEW ONE. BASED ON THE MANUFACTURING DOCUMENTATION REVIEW, THERE IS NO INDICATION THAT THE EVENT IS RELATED TO THE DEVICE MANUFACTURING PROCESS. AS PART OF THE POST MARKET SURVEILLANCE PROGRAM, INFORMATION FROM THIS COMPLAINT IS TRENDED FOR STATISTICAL SIGNALS AND CORRECTIVE / PREVENTIVE ACTION MAY BE TRIGGERED AT A LATER TIME. SINCE THERE WAS NO EVIDENCE TO SUGGEST THE EVENT WAS RELATED TO A MANUFACTURING OR DESIGN ISSUE, NO CORRECTIVE ACTIONS WILL BE TAKEN AT THIS TIME. UPDATED SECTIONS: B.4, G.3, G.6. H.2, H.3, H.6, AND H.11. THE MANUFACTURER WILL SUBMIT A SUPPLEMENTAL REPORT IF NEW FACTS ARISE WHICH MATERIALLY ALTER INFORMATION SUBMITTED IN A PREVIOUS MDR REPORT. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN 30 DAYS OF RECEIPT.
MANUFACTURER¿S REF. NO: (B)(4). THE PURPOSE OF THIS MDR SUBMISSION IS TO REPORT THAT THE PRODUCT ANALYSIS LAB RECEIVED THE COMPLAINT DEVICE ON 05-FEB-2025. A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED ONCE THE PRODUCT INVESTIGATION HAS BEEN COMPLETED. THE MANUFACTURER WILL SUBMIT A SUPPLEMENTAL REPORT IF NEW FACTS ARISE WHICH MATERIALLY ALTER INFORMATION SUBMITTED IN A PREVIOUS MDR REPORT. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN 30 DAYS OF RECEIPT.
MANUFACTURER¿S REF. NO: (B)(4). INFORMATION REGARDING PATIENT IDENTIFIER, DATE OF BIRTH, AGE, SEX, GENDER, WEIGHT, RACE, AND ETHNICITY WERE NOT PROVIDED. SECTION E.1: THE INITIAL REPORTER PHONE: (B)(6). SECTION H.3: THE DEVICE IS AVAILABLE TO BE RETURNED FOR EVALUATION AND TESTING. HOWEVER, IT HAS NOT BEEN RECEIVED TO DATE. IF THE DEVICE RETURNS, A DEVICE INVESTIGATION WILL BE PERFORMED. LAKE REGION MEDICAL PERFORMED A REVIEW OF THE DEVICE HISTORY RECORDS RELATIVE TO THE MANUFACTURING, INSPECTION, AND PACKAGING OF THE LOT 9080733. THE HISTORY RECORD INDICATES THIS PRODUCT WAS FINAL INSPECTION TESTED AT LAKE REGION MEDICAL AND WAS DETERMINED TO BE ACCEPTABLE. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY JOHNSON & JOHNSON MEDTECH, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, JOHNSON & JOHNSON MEDTECH, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. MISSING INFORMATION FROM THIS REPORT IS IDENTIFIED AS BLANK; THIS INFORMATION WAS NOT PROVIDED IN THE REPORTED EVENT OR AVAILABLE AT THE TIME OF REPORT SUBMISSION. THE MANUFACTURER WILL SUBMIT A SUPPLEMENTAL REPORT IF NEW FACTS ARISE WHICH MATERIALLY ALTER INFORMATION SUBMITTED IN A PREVIOUS MDR REPORT. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN 30 DAYS OF RECEIPT.
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION OF A RIGHT INTERNAL CAROTID CAVERNOUS SINUS ANEURYSM, THE 4.5MM X 22MM ENTERPRISE® VASCULAR RECONSTRUCTION DEVICE (ENC452212 / 9080733) WAS IMPEDED IN THE HUB OF THE CONCOMITANT MICROCATHETER (UNSPECIFIED BRAND) AND COULD NOT BE FURTHER ADVANCED. THE PHYSICIAN RETRACTED ONLY THE STENT, BUT IT WAS OBSERVED TO HAVE DETACHED FROM THE DELIVERY WIRE IN THE Y-CONNECTOR. THE PHYSICIAN REMOVED THE Y-CONNECTOR AND THE STENT AND REPLACED THE STENT TO COMPLETE THE PROCEDURE USING THE SAME MICROCATHETER. THERE WAS NO REPORT OF ANY NEGATIVE PATIENT IMPACT. ON 13-JAN-2025, ADDITIONAL INFORMATION WAS RECEIVED. THE INFORMATION CONFIRMED THE PROCEDURE WAS EMBOLIZATION OF THE RIGHT INTERNAL CAROTID CAVERNOUS SINUS ANEURYSM. A CONTINUOUS FLUSH HAD BEEN MAINTAINED THROUGH THE MICROCATHETER (UNSPECIFIED BRAND). THE STENT WAS SEPARATED FROM THE PUSH GUIDE WIRE WHEN IT WAS REMOVED. THERE WAS DAMAGE NOTED ON THE MICROCATHETER, HOWEVER, THE INFORMATION DID NOT DESCRIBE THE DAMAGE. THE REPLACEMENT DEVICE WAS NOT ANOTHER 4.5MM X 22MM ENTERPRISE® VASCULAR RECONSTRUCTION DEVICE (ENC452212).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1501589 | CERENOVUS ENTERPRISE | INTRACRANIAL NEUROVASCULAR STENT | NJE | MEDOS INTERNATIONAL SARL | 9080733 | 10886704043980 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | MICROCATHETER (UNSPECIFIED BRAND) |