FDA Adverse Event Malfunction Summary report: N

HEARTSTART MRX -EMS DEFIBRILLATOR

MDR report key: 2123540 · Received June 9, 2011

Report

Report Number
1218950-2011-01654
Event Type
Malfunction
Date Received
June 9, 2011
Report Date
May 17, 2011
Manufacturer
PHILIPS HEALTHCARE
Product Code
MKJ
PMA / PMN Number
K031187
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE CUSTOMER REPORTED A FAILURE TO SHOCK. THERE WAS NO REPORT OF PT INVOLVEMENT. THE DEVICE WAS EVALUATED BY THE CUSTOMER'S BIOMED. THERE WERE CHARGE/SHOCK FAILURE/THERAPY PCA MESSAGES IN THE STATUS LOG. THE THERAPY PCA WAS REPLACED TO RESOLVE THE ISSUE. THE DEVICE PASSED ALL TESTING AND WAS RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A FAILURE TO SHOCK. THERE WAS NO REPORT OF PT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEARTSTART MRX -EMS DEFIBRILLATOR MKJ PHILIPS HEALTHCARE M3536A

Patients

Seq Age Sex Outcome Treatment
1