FDA Adverse Event
Malfunction
Summary report: N
HEARTSTART MRX -EMS DEFIBRILLATOR
MDR report key: 2123540
·
Received June 9, 2011
Report
- Report Number
- 1218950-2011-01654
- Event Type
- Malfunction
- Date Received
- June 9, 2011
- Report Date
- May 17, 2011
- Manufacturer
- PHILIPS HEALTHCARE
- Product Code
- MKJ
- PMA / PMN Number
- K031187
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE CUSTOMER REPORTED A FAILURE TO SHOCK. THERE WAS NO REPORT OF PT INVOLVEMENT. THE DEVICE WAS EVALUATED BY THE CUSTOMER'S BIOMED. THERE WERE CHARGE/SHOCK FAILURE/THERAPY PCA MESSAGES IN THE STATUS LOG. THE THERAPY PCA WAS REPLACED TO RESOLVE THE ISSUE. THE DEVICE PASSED ALL TESTING AND WAS RETURNED TO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED A FAILURE TO SHOCK. THERE WAS NO REPORT OF PT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HEARTSTART MRX -EMS DEFIBRILLATOR | MKJ | PHILIPS HEALTHCARE | M3536A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |