FDA Adverse Event
Malfunction
Summary report: N
SPRINT QUATTRO SECURE S
MDR report key: 2123525
·
Received June 14, 2011
Report
- Report Number
- 2649622-2011-07976
- Event Type
- Malfunction
- Date Received
- June 14, 2011
- Date of Event
- March 2, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S039
- Removal / Correction Number
- Z-0474-2011
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE LEAD IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY FOR (B)(4): HELIX DISENGAGED FROM HELICAL CHANNEL. FULL LEAD RETURNED AND ANALYZED. ADDITIONAL FINDING OF BLOOD IN/ON HELIX MECHANISM.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING THE IMPLANT ATTEMPT THE HELIX WOULD NOT EXTEND ON THE RIGHT VENTRICULAR LEAD. THE LEAD WAS REMOVED AND A NEW LEAD WAS USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE S | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6935 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Other |