FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 3000 SYSTEM
MDR report key: 2123518
·
Received June 8, 2011
Report
- Report Number
- 3008642652-2011-00023
- Event Type
- Malfunction
- Date Received
- June 8, 2011
- Date of Event
- May 5, 2011
- Report Date
- June 8, 2011
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL SUMMARY: DEVICE EVAL OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (BENT PINS/DAMAGED E-BELT CONNECTOR) HAS BEEN CONFIRMED. UPON INVESTIGATION, PINS IN THE ELECTRODE BELT'S TRUNK CABLE CONNECTOR WERE BENT. THE CAUSE FOR THE BENT PINS CANNOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY THE RESULT OF EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE ELECTRODE BELT CONNECTOR. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.
Description of Event or Problem · 1
A (B)(6) MALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT HE COULD NOT CONNECT HIS MONITOR TO HIS BELT. THE PT WAS PROVIDED WITH A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 3000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 3000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR |