ALARIS PUMP MODULE
Report
- Report Number
- 2016493-2011-00367
- Event Type
- Malfunction
- Date Received
- June 8, 2011
- Date of Event
- February 28, 2011
- Report Date
- March 7, 2011
- Manufacturer
- CAREFUSION CORP.
- Product Code
- FRN
- PMA / PMN Number
- K012383
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHARMACIST
Narratives
MFR'S REPORT DATE: 06/08/2011. (B)(4). THE CUSTOMER'S REPORT OF AN OVER INFUSION OF PROPOFOL COULD NOT BE DEFINITIVELY CONFIRMED NOR REPLICATED DURING THE INVESTIGATION. A REVIEW OF THE ASSOCIATED PC UNIT EVENT LOGS FOUND THAT THE DEVICE WAS PROGRAMMED TO INFUSE PROPOFOL WITH THE DOSAGE BEING TITRATED DOWN THROUGHOUT THE TIME PERIOD IN QUESTION WITH THE INFUSION ENDING IN AN AIR-IN-LINE ALARM WITH 26 MLS OF THE STARTING 50 MLS UNACCOUNTED FOR. VISUAL INSPECTION OF THE DEVICE NOTED AN ISSUE THAT COULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE PLATEN POST WAS FOUND TO HAVE BEEN BROKEN OFF COMPLETELY. PRIOR INVESTIGATIONS HAVE REVEALED A POTENTIAL FOR CYCLICAL PERIODS OF UNREGULATED FLOW WITH A BROKEN PLATEN POST ON THE PLATEN ASSEMBLY. ADD'L DAMAGE OBSERVED TO THE BEZEL OF THE DEVICE SUGGESTS A FOREIGN OBJECT MAY HAVE BEEN CAUGHT BETWEEN THE DOOR AND PLATEN LEADING TO THE PLATEN POST DAMAGE. TESTING FAILED TO INDUCE THE UNREGULATED FLOW SCENARIO. THE RESULTS OF SEVERAL RATE ACCURACY TESTS ON THE DEVICE, WITH THE DAMAGED PLATEN AND WITH A KNOWN GOOD PLATEN, INDICATE THAT THE DEVICE IS DELIVERING FLUID SLIGHTLY OUT OF THE +/- 5% SPECIFICATIONS. TESTING ALSO INDICATES THAT THE BROKEN PLATEN HAD NO CONTRIBUTORY EFFECT ON THE OUT OF SPECIFICATION CONDITION. THE ROOT CAUSE OF THE CUSTOMER'S REPORTED PROBLEM WAS NOT DETERMINED.
CUSTOMER REPORTED AN OVER INFUSION OF PROPOFOL. PROPOFOL 20 MCG/KG/MIN (1000MG/100ML) INITIATED AT 1543, AND FOUND TO BE COMPLETELY EMPTY AT 1714. WHEN THE PUMP WAS TOTALLED OUT, THE VOLUME INFUSED SHOWED 75ML (WHICH WAS THE VOLUME FOR A TOTAL OF 4.5 HOURS INCLUDING THE AM). WHEN THE RN CHECKED THE CHANNEL AND PAUSED IT, THE DRIP CHAMBER CONTINUED TO VERY STEADILY DRIP. IT IS ASSUMED THAT THE PT RECEIVED THE ENTIRE BOTTLE OF PROPOFOL. THERE WAS NO INJURY TO THE PT AND NO MEDICAL INTERVENTION WAS REQUIRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ALARIS PUMP MODULE | FRN | CAREFUSION CORP. | 8100 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 37 YR | ALARIS PC UNIT, S/N (B)(4)| ALARIS PUMP MODULE, S/N (B)(4)| ALARIS PUMP MODULE, S/N (B)(4)| ALARIS PUMP MODULE, S/N (B)(4) |