FDA Adverse Event Malfunction Summary report: N

PETITE PACING LEAD

MDR report key: 2123484 · Received June 7, 2011

Report

Report Number
1035166-2011-00020
Event Type
Malfunction
Date Received
June 7, 2011
Date of Event
March 24, 2011
Report Date
June 3, 2011
Manufacturer
OSCOR INC.
Product Code
DTB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE LEAD WAS SURGICALLY ABANDONED (CAPPED) WITH NO ADVERSE PATIENT EFFECTS REPORTED, THEREFORE IT WILL NOT BE RETURNED FOR ANALYSIS. AS A RESULT, THE ALLEGATIONS AGAINST THIS LEAD CANNOT BE CONFIRMED. THE EVENT WILL BE RE-EVALUATED IF ADDITIONAL INFORMATION BECOMES AVAILABLE. LOSS OF CAPTURE/SENSING IS A RECOGNIZED CLINICAL EVENT REFERENCED IN THE DEVICE'S LABELING AND/OR REPORTED IN THE MEDICAL LITERATURE. THE DEVICE HISTORY RECORDS AND THE COMPLAINT FILES OF THE LEAD MODEL WERE REVIEWED. NO TREND OF THE SAME OR SIMILAR TYPE WAS FOUND OR INDICATED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED ON (B)(4) 2011, "THE LEADS WERE CAPPED DUE TO LOSS OF VENTRICULAR CAPTURE AND PACING WHICH LED TO 9 SECOND PAUSES. THE CUSTOMERS PACEMAKER WAS ALSO EXPLANTED. NO ADVERSE PATIENT EFFECTS WERE REPORTED. ON (B)(4) 2011, THE CUSTOMER REVISED THEIR REPORT TO THE FOLLOWING: "ACCORDING TO THE HOSPITAL RECORDS AND THE SALES REPRESENTATIVE, THE ATRIAL LEAD HAD DISLODGED SOME TIME AGO AND THE PATIENT WAS IN A-FIB AT THAT TIME SO THE DEVICE WAS REPROGRAMMED TO VVI. THE PATIENT WAS ADMITTED ON (B)(6) 2011, WITH THE HOPE TO CONVERT TO SINUS RHYTHM AND REVISE THE ATRIAL LEAD. IT WAS THEN FOUND THAT THE VENTRICULAR LEAD WAS NOT CAPTURING OR PACING WHICH LED TO 9 SECOND PAUSES. BOTH LEADS WERE CAPPED AND THE CUSTOMER'S PACEMAKER WAS EXPLANTED. IT WAS REPORTED THAT THE INTERROGATION THAT WAS PERFORMED FOUND THE RV LEAD HAD MALFUNCTIONED. NO ADVERSE PATIENT EFFECTS WERE REPORTED." THE ATRIAL LEAD WAS ACTIVELY IMPLANTED FOR APPROXIMATELY 4 YEARS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PETITE PACING LEAD PASSIVE FIXATION PACING LEAD DTB OSCOR INC. PETITE 52 RB 6391

Patients

Seq Age Sex Outcome Treatment
1 73 YR