FDA Adverse Event Injury Summary report: N

SPRINT FIDELIS

MDR report key: 2123459 · Received June 14, 2011

Report

Report Number
2649622-2011-08127
Event Type
Injury
Date Received
June 14, 2011
Date of Event
March 7, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S029
Removal / Correction Number
Z-0067-0070-2008
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE SPRINT FIDELIS LEAD MODEL IS INCLUDED IN A FIELD ADVISORY. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. IT WAS ALSO NOTED THAT THE OUTER INSULATION WAS BREACHED CUT, THE OUTER INSULATION HAD A COSMETIC DEPRESSION, THERE WAS BLOOD IN/ON THE HELIX MECHANISM, THERE WAS APPARENT EXPLANT DAMAGE AND THERE WAS WHITE SUBSTANCE ON THE LEAD. (B)(4) THE PARTIAL LEAD IN SEGMENTS WAS RETURNED, ANALYZED, AND THERE WAS INTERMITTENCY BETWEEN THE PIN AND THE CAP IN THE CONNECTOR. IT WAS ALSO NOTED THAT THE DISTAL AND DEFIBRILLATOR CONDUCTORS WERE DISTORTED, SEVERAL CONDUCTORS HAD BLOOD/BODY FLUID (NOT OBSTRUCTED), THERE WAS BLOOD/BODY FLUID IN THE OUTER TUBING OVERLAY, THE INNER INSULATION WAS KINKED/BUCKLED, THE OUTER TUBING OVERLAY WAS MELTED, THE OUTER TUBING OVERLAY HAD COSMETIC ENVIRONMENTAL STRESS CRACKING AND A NON-ELECTRICAL BREACH, THE OUTER INSULATION HAD A COSMETIC DEPRESSION, AND THERE WAS BLOOD IN/ON THE HELIX MECHANISM.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PROPHYLACTIC EXTRACTION OF THE RIGHT VENTRICULAR LEAD, THE ATRIAL LEAD WAS DISLODGED. BOTH LEADS WERE EXPLANTED AND REPLACED. THE RIGHT VENTRICULAR LEAD TESTED OUT OF SPECIFICATION DURING ANALYSIS. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT FIDELIS IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6949 ASKU

Patients

Seq Age Sex Outcome Treatment
1 71 YR Hospitalization| R 7288 IMPLANTABLE PACEMAKER/CARDIO/DEFIB