FDA Adverse Event Injury Summary report: N

ENRHYTHM DR

MDR report key: 2123438 · Received June 14, 2011

Report

Report Number
6000144-2011-02269
Event Type
Injury
Date Received
June 14, 2011
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
DXY
PMA / PMN Number
P980035/S38
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY (B)(4) THE DEVICE WAS RETURNED AND ANALYSIS FOUND HIGH BATTERY IMPEDANCE. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE PREMATURELY REACHED ELECTIVE REPLACEMENT INDICATOR (ERI). THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENRHYTHM DR IMPLANTABLE PULSE GENERATOR DXY MEDTRONIC MED REL, INC. P1501DR ASKU

Patients

Seq Age Sex Outcome Treatment
1 87 YR Hospitalization| R 4074 IMPLANTABLE PACING LEAD| 4076 IMPLANTABLE PACING LEAD