ATTAIN OTW
MDR report key: 2123425
·
Received June 14, 2011
Report
- Report Number
- 2649622-2011-08122
- Event Type
- Injury
- Date Received
- June 14, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P010015/S012
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR (B)(4): NO ANOMALIES FOUND. FULL LEAD RETURNED AND ANALYZED. ADDITIONAL FINDING OF ALL CONDUCTORS BLOOD/BODY FLUID (NOT OBSTRUCTED).
Description of Event or Problem · 1
IT WAS REPORTED THAT THE LEFT VENTRICULAR LEAD EXHIBITED HIGH THRESHOLDS. DURING THE REPLACEMENT PROCEDURE, A SECOND LEFT VENTRICULAR LEAD IMPLANTATION WAS ATTEMPTED, BUT NOT SUCCESSFUL BECAUSE THE LEAD COULD NOT ACHIEVE ADEQUATE PLACEMENT. THE LEAD WAS REMOVED AND NOT REPLACED. THE CHRONIC LEFT VENTRICULAR LEAD IS STILL IN USE. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ATTAIN OTW | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 4194 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 45 YR | Hospitalization| R | (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| (B)(4) IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE TACHY LEAD |