FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 2123384
·
Received June 14, 2011
Report
- Report Number
- 2531779-2011-04139
- Event Type
- Malfunction
- Date Received
- June 14, 2011
- Report Date
- May 17, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS AND EVALUATED. THE PUMP WAS INTERMITTENTLY UNRESPONSIVE TO ALL KEYPAD BUTTON PRESSES. THE KEYPAD WAS REMOVED AND ADHESIVE WAS FOUND UNDER THE BUTTON CONTACTS. THE BOLUS BUTTON WAS ALSO TORN.
Description of Event or Problem · 1
ON (B)(6) 2011, THE LAY-USER/PATIENT CONTACTED ANIMAS ALLEGING THAT KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE PATIENT CLAIMED THAT THE PUMP WAS INTERMITTENTLY UNRESPONSIVE TO BUTTON PRESSES. SHE ALSO CLAIMED THAT THE BUTTONS WERE STICKING. THE PATIENT DID NOT ALLEGE ANY HARM OR INJURY BECAUSE OF THE REPORTED ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 37 YR |