FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2123369 · Received June 14, 2011

Report

Report Number
2649622-2011-07957
Event Type
Injury
Date Received
June 14, 2011
Date of Event
March 1, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM.

Description of Event or Problem · 1

IT WAS REPORTED THAT FOLLOWING THE IMPLANT, THE PATIENT WENT TO THE HOSPITAL COMPLAINING OF CHEST PAINS ASSOCIATED WITH RIGHT VENTRICULAR (RV) PACING. IT WAS DETERMINED THAT THE RV LEAD HAD PERFORATED THE RV AND THE PATIENT HAD PERICARDIAL EFFUSION. THE LEAD WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 83 YR Hospitalization| L| R ADDR01 IMPLANTABLE PACEMAKER| 5568 IMPLANTABLE PACING LEAD