FDA Adverse Event
Malfunction
Summary report: N
PARADIGM INSULIN INFUSION PUMP
MDR report key: 2123359
·
Received June 10, 2011
Report
- Report Number
- 3004209178-2011-81727
- Event Type
- Malfunction
- Date Received
- June 10, 2011
- Date of Event
- March 23, 2011
- Report Date
- June 1, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS MED-REL
- Product Code
- LZG
- PMA / PMN Number
- K040676
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TU
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP FAILED THE DISPLACEMENT TEST AND A MOTOR ERROR ALARM OCCURRED DURING REWIND DUE TO AN OUT-OF-PHASE MOTOR ENCODER SIGNAL.
Description of Event or Problem · 1
THE CUSTOMER REPORTED RECEIVING THREE MOTOR ERROR ALARMS. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | MEDTRONIC PUERTO RICO OPERATIONS MED-REL | MMT-715WWL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |