INRATIO
Report
- Report Number
- 2027969-2011-01309
- Event Type
- Malfunction
- Date Received
- June 10, 2011
- Date of Event
- May 13, 2011
- Report Date
- June 10, 2011
- Manufacturer
- ALERE SAN DIEGO, INC.
- Product Code
- GJS
- PMA / PMN Number
- K021923
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MT, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED: DATE: (B)(6) 2011, INRATIO METER = 5.0 INR, REFERENCE = 3.3 INR , MEAN = 4.15, CONFIDENCE LIMITS = 2.4-6.1. THE MEAN OF THE INRATIO METER AND COMPARATIVE SYSTEM INR WERE CALCULATED. BOTH INRATIO AND REFERENCE VALUES ARE WITHIN THE CONFIDENCE LIMITS FOR INR TESTING. THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. INR REPORTED HAVE MET THE CRITERIA FOR ACCURACY. AS OF 06/09/2011, NO STRIP LOT INFO WAS PROVIDED NOR IS PRODUCT EXPECTED TO RETURN. NO FURTHER INVESTIGATION WILL BE PURSUED AT THIS TIME. CONCLUSION: ANALYSIS OF THE CLIENT'S DATA FROM INRATIO AND REFERENCE TESTS REVEALED THAT TEST RESULT COMPARISON MET ACCURACY CRITERIA. NO STRIP LOT INFO WAS PROVIDED, AND NO PRODUCT WAS EXPECTED TO BE RETURNED. ROOT CAUSE COULD NOT BE DETERMINED WITH THE INFO CUSTOMER PROVIDED. STRIPS WERE BEING STORED IN A CAR. PER PRODUCT USER GUIDE - STORAGE AND HANDLING INSTRUCTION, "STORE STRIPS AT ROOM TEMP (BELOW 90 DEGREES FAHRENHEIT) UNTIL EXPIRED." THIS ISSUE WILL BE SUBJECT TO TRACKING AND TRENDING. CORRECTIVE ACTION NOT REQUIRED AT THIS TIME.
CALLER ALLEGED DISCREPANT RESULTS USING THE INRATIO METER. RESULTS AS FOLLOWS: DATE: (B)(6) 2011, INRATIO: 5.0, LAB: 3.3. TESTS WERE DONE WITHIN AN HOUR OF EACH OTHER. LAB DRAW WAS DONE DUE TO PT'S HIGH INR RESULT. NO UNUSUAL BRUISING/BLEEDING WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INRATIO | PROTHROMBIN TIME TEST | GJS | ALERE SAN DIEGO, INC. | 100139 | NG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |