FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 2123342
·
Received June 10, 2011
Report
- Report Number
- 2032227-2011-01430
- Event Type
- Malfunction
- Date Received
- June 10, 2011
- Date of Event
- May 29, 2011
- Report Date
- May 29, 2011
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP FAILED THE DISPLACEMENT TEST AND ALARMED MOTOR ERROR DURING THE REWIND DUE TO AN OUT OF PHASE MOTOR ENCODER SIGNAL.
Description of Event or Problem · 1
THE CUSTOMER CALLED TO REPORT A MOTOR ERROR ALARM. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP FAILED THE DISPLACEMENT TEST. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | MDS | MEDTRONIC MINIMED | MMT-522LNAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR |